ResMed 2011-09, 248551/1 manual Other troubleshooting

Models: 2011-09 248551/1

1 38
Download 38 pages 47.37 Kb
Page 29
Image 29
Note: The alarm will be cleared automatically after one minute.

Problem / possible cause

Action

 

 

LCD: Low FiO2 Level!

FiO2 has dropped below the pre-set alarm level.

Perform FiO2 sensor calibration.

If the problem persists contact your clinician.

LCD: High FiO2 Level!

FiO2 has exceeded the pre-set alarm

• Perform FiO2 sensor calibration.

level.

• If the problem persists contact your

 

clinician.

 

 

LCD: Keypad Failure!

 

 

 

One of the keys was held down for more

Remove any blockages from the keypad.

than 10 seconds or got stuck.

 

LCD: Attention High Temp [42, 43, 44, 45]!

The temperature inside the device is high.

Ensure the ambient temperature conditions are within the specified operating range.

LCD: Internal Battery Use!

The device is using the internal battery.

Check if the power cord is properly connected to the device if you want to run from mains power.

Press the Alarm mute button Other troubleshootingManual background to cancel the alarm.

LCD: External DC Power Use!

The device is powered by an external battery.

Check if the AC power cord is properly connected to the device if you want to run from mains power.

Note: The alarm will be cleared automatically after one minute.

Other troubleshooting

Problem / possible cause

Solution

 

 

No display

Power failure. The device stops delivering air pressure.

Remove the mask or the catheter mount from the tracheostomy tube until power is restored.

Power not connected or device is not switched on.

Ensure the power cord is connected and press the switch at the back of the device once.

Treatment pressure seems low

Ramp time is in use.

Wait for air pressure to build up.

Air filter is dirty.

Replace air filter.

 

 

Air tubing is kinked or punctured.

Straighten or replace tubing.

 

 

Air tubing is not connected properly.

Connect the air tubing firmly at both

 

ends.

Mask and headgear not positioned correctly.

Adjust position of mask and headgear.

Plug(s) missing from access port(s) on

Replace plug(s).

mask.

 

 

 

24

Page 29
Image 29
ResMed 2011-09, 248551/1 manual Other troubleshooting, Note The alarm will be cleared automatically after one minute