TECHNICAL SPECIFICATIONS

TECHNICAL SPECIFICATIONS

PERFORMANCE

Maximum heater plate temperature: 185°F (85°C)

Temperature cut-out: 199°F (93°C)

Maximum output humidity: 95% relative humidity (control dial set to maximum setting, flow rate 50 L/min, ambient temperature of 68°F [20°C], ambient relative humidity of 40%).

Maximum gas temperature: 106°F (41°C)

DIMENSIONS (H X W X D)

Docking station and water chamber: 4.6" x 6.7" x 8.3" (118 mm x 170 mm x 212 mm)

WEIGHT

Docking station and unfilled water chamber: 1.5 lb. (670 g)

Water capacity to maximum fill line: 415 mL

POWER SUPPLY

Input range for ResMed S8 flow generator with HUMIDAIRE 3i: 100–240V, 50/60Hz; 110V, 400Hz; 2.5A <140 VA (110 W) (maximum power consumption). Instantaneous peak power consumption <340 VA.

Actual average power consumption will vary, depending on factors such as temperature setting, set pressure, the use of accessories, height above sea level and ambient temperature.

Maximum heater element power: 85 W.

HOUSING CONSTRUCTION

Docking station: Flame retardant engineering thermoplastic, plated cast aluminium.

Water chamber: Injection molded plastic, stainless steel.

ENVIRONMENTAL CONDITIONS

Operating temperature: +41°F to +104°F (+5°C to +40°C)

Operating humidity: 10–95% non-condensing

Storage and transport temperature: -4°F to +140°F (-20°C to +60°C)

Storage and transport humidity: 10–95% non-condensing

IEC 60601-1 CLASSIFICATION Class II (double insulation) Type CF

ELECTROMAGNETIC COMPATIBILITY

Please refer to your flow generator user's manual for details.

Note: The manufacturer reserves the right to change these specifications without notice.