Siemens 1015N-2MFM-1A manual Communications Configuration, WINDOWS HyperTerminalTM SETUP

Models: 1015N-2MFM-1A

1 25
Download 25 pages 49.16 Kb
Page 13
Image 13
6.COMMUNICATIONS CONFIGURATION

1015N-2MFM-1A

6.COMMUNICATIONS CONFIGURATION

The 1015N-2M Modem Card has been pre-configured at the factory to allow for proper interface with Controlotron’s System 1010N flowmeters.

CAUTION: The 1015N-2M Modem Card will not function properly if the precon- figured settings are altered. If the settings are altered, refer to reconfiguration procedure below.

7.WINDOWS HyperTerminalTM SETUP

Step 1. Invoke HyperTerminalTM (generally found in the Program Files / Accessories Folder).

Step 2. Give your connection a name in the space provided.

Step 3. Type in your meter’s telephone number.

Step 4. Then click on Modify, then Configure and select 2400 baud as the modem speed.

Step 5. Now click on Advanced Tab and set Data Bits to 7 and Parity to Odd. You may leave Stop Bits at 1.

Step 6. Then click on Settings, Terminal Setup and pick VT-100 as your preferred terminal emulation. Then click OK.

Step 7. Now click on Dial and your modem should dial up the flowmeter.

Step 8. Once your modem answers (you should hear the carrier tones while the modems are negotiating), click on ‘?’ to see the list of available direct execute commands from the meter.

Step 9. If you wish to use the meter menu, just type: Menu.

Step 10. You can direct HyperTerminalTM to save the contents of your session to a named file by clicking on Transfer and then Capture Text. You may use the default file name to capture the flow meter data, which may then be copied to another file using your text editor.

Step 11. To hang up, simply terminate HyperTerminalTM and the remote modem will hang up automatically and ready itself for the next call.

1015N-2M Reconfiguration Procedure

To reconfigure the 1015N-2M Modem Card to factory settings (see Appendix A for commands), proceed as follows:

1.Set up HyperTerminalTM as in the first six steps of Section 7 above.

2.Type: AT &F &W0 &W1 and press <ENTER> key. (Modem prints out: OK.)

3.Type: AT E0 M0 Q1 &C1 S0=1 &W0 &W1 and press <ENTER> key.

4.Type: AT &V and press <ENTER> key. Verify on the terminal that the following com- mands are individually stored in:

Active Profile: E0 M0 Q1 &C1 S01=1 &W0 &W1

Stored Profile 0: E0 M0 Q1 &C1 S01=1 &W0 &W1

Stored Profile 1: E0 M0 Q1 &C1 S01=1 &W0 &W1

5.Complete steps 7 through 11 of Section 7 above.

9

Page 13
Image 13
Siemens 1015N-2MFM-1A manual Communications Configuration, WINDOWS HyperTerminalTM SETUP