53
HeartView CT
Calcium Scoring evaluation is performed on a separate
syngotask card:
1. The threshold of 130 HU is applied for score
calculation by default, however, you can modify it
accordingly.
2. In addition to the seeding method, you can use
freehand ROI for the definition of lesions.
3. The separation and modification of lesions within
a defined volume (depth in mm) can be performed
not only on 2D slices, but also with 3D editing.
4. For easier identification of small lesions, you can
blowup the display.
5. You can customize hospital/office information on
the final report using Report Configuration.
6. You can generate HTML report including site spe-
cific information, free text and clinical images. This
then can be saved on floppy disc and/or printed.
7. The results are displayed online in a separate
segment including the following information:
– Area (in mm3)
– Peak density (in HU)
– Volume (in mm3)
– Calcium mass (mg calcium Hydroxyapatite)
– Score (Agatston method)
8. The results can be printed on laser film, paper
printer or saved into data base.
User interface of syngo Calcium Scoring