87
Children
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1–2 ml per kg of body weight
Tips
• Children with a body weight of more than 54 kg
should be examined with an adult protocol.
• The first recon job is defined for image viewing and
filming.
MIPthin: AbdCTA sagittal
Image thickness 3
Distance between images 3
Number of images 60
For the 2ndreconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a sagittal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.