RS232 SERIAL COMMUNICATIONS

(Special Order Only)

����������������� �

��������

[Z] [CR] [LF]

 

Zero the displayed reading

[F] [CR] [LF]

 

Calibrate Full Scale (Input signal

 

 

is scaled to display Full Scale)

[S] [counts] [CR] [LF]

 

Set Full Scale display value.

 

 

(counts range = 0 .... 99999)

[D] [value] [CR] [LF]

 

Set decimal position

 

 

1

= x.xxxx

 

 

2

= xx.xxx

 

 

3

= xxx.xx

 

 

4

= xxxx.x

 

 

5

= xxxxx

[P1] [value] [CR] [LF]

 

Set level of Lo Alarm #1

[P2] [value] [CR] [LF]

 

Set level of Lo Alarm #2

[P3] [value] [CR] [LF]

 

Set level of Lo Alarm #3

[P4] [value] [CR] [LF]

 

Set level of Lo Alarm #4

[H1] [counts] [CR] [LF]

 

Set Lo Alarms #1&2 hysteresis

[H2] [counts] [CR] [LF]

 

Set Hi Alarms #1&2 hysteresis

�����������������

���������

��������������

[R5] [CR] [LF]

 

Pressure value

 

[P] [value] [CR] [LF]

[R8] [CR] [LF]

 

Full Scale display

[S] [counts] [CR] [LF]

[R9] [CR] [LF]

 

Decimal location

[D] [value] [CR] [LF]

 

 

 

 

1

= x.xxxx

 

 

 

 

2

= xx.xxx

 

 

 

 

3

= xxx.xx

 

 

 

 

4

= xxxx.x

 

 

 

 

5

= xxxxx

[R1] [CR] [LF]

 

Lo Alarm #1 level

[P1] [value] [CR] [LF]

[R2] [CR] [LF]

 

Hi Alarm #1 level

[P2] [value] [CR] [LF]

[R3] [CR] [LF]

 

Lo Alarm #2 level

[P3] [value] [CR] [LF]

[R4] [CR] [LF]

 

Hi Alarm #2 level

[P4] [value] [CR] [LF]

[R6] [CR] [LF]

Lo Alarm #1&2 hysteresis

[H1] [counts] [CR] [LF]

[R7] [CR] [LF]

Hi Alarm #1&2 hysteresis

[H2] [counts] [CR] [LF]

[RX] [CR] [LF]

 

Setpoint value

 

[X] [counts] [CR] [LF]

SERIAL LOCKOUT:

The Model 951 will respond to RS-232 commands and queries while it is in the normal display mode. When it is in the menu entry or calibration modes

it will not respond to RS-232 commands or queries.

Page 16

Page 17
Image 17
Sierra 951 instruction manual Serial Lockout