Owner’s Manual - Oxygen Therapy Door

Removal - New-style Gauge

1.Open the cage door.

2.With a Phillips screwdriver and a 3/8 in. wrench, remove the two gauge bracket screws/nuts (Figure 24).

3.Remove the gauge bracket and the gauge from the inside of the oxygen door.

Gauge

Bracket

Gauge Bracket Screws

Gauge

Bracket

Figure 24. Front View of the New-style Gauge (mounted on door)

Figure 25. Rear View of the New-style Gauge (removed from door)

Installation of New-style Gauge

1.Order the following from SSCI:

„

New-style Gauge -

P/N 854771

„

Gauge Bracket -

P/N 600162-2

Note: Order a new gauge bracket only if replacing an old-style gauge. If replacing a new-style gauge, use your existing bracket.

2.Loosely mount the gauge bracket (Figure 25) onto the inside of the oxygen door with the screws and nuts removed above. Use the existing gauge bracket mounting holes.

3.Place the gauge into the gauge bracket making sure the gauge is right-side-up.

4.Tighten the bracket screws and nuts to secure the gauge.

5.On the front of the polycarbonate door, mark the centers of the two buttons on the face of the gauge (Figure 12).

Form No. 702703 Rev. E August, 2006

23

Page 27
Image 27
Suburban Mfg 12155-00-EPEPAA, 12155-00-FNHJAA Removal New-style Gauge, Front View of the New-style Gauge mounted on door