Performance Tests
TDS 684A, TDS 744A, & TDS 784A Service Manual 4–29
Align the active cursor
over the waveform. 1
Read the
measurement
results. 2
Figure 4–4: Measurement of DC Offset Accuracy at Zero Setting
g. Test all channels: Repeat substeps a through f for all input channels.
3. Disconnect the hookup: No hookup was required.
WARNING. The generator is capable of outputting dangerous voltages. Be sure to
set the DC calibration generator to 0 volts before connecting, disconnecting,
and/or moving the test hookup during the performance of this procedure.
Equipment
Required Two dual-banana connectors (Item 6)
One BNC T connector (Item 7)
One DC calibration generator (Item 9)
Two precision coaxial cables (Item 5)
Prerequisites The oscilloscope must meet the prerequisites listed on page 4–15
Check DC VoltageMeasurement Accuracy