Warranted Characteristics
1–20 TDS 684A, TDS 744A, & TDS 784A Service ManualTable 1–14: Warranted Characteristics — Environmental (Cont.)
Name Description
Susceptibility1, 2Meets or exceeds the EMC requirements of the following standards:
EN 50082–1 European Community Requirements
IEC 801-2 Electrostatic Discharge
IEC 801-3 Radiated Susceptibility
IEC 801-4 Fast Transients
IEC 801-5 AC Surge
Third Party Certification Conforms to and is certified where appropriate to:
UL 1244
CSA–C22.2 No. 231
1VGA output cable needs to be terminated, if connected at all, for the Instrument to meet these standards. The test will
pass with LCOM part # CTL3VGAMM–5.
2The GPIB cable connected to the instrument for certain of the emissions tests must be “low EMI” having a high-quality
outer shield connected through a low impedance to both connector housings. Acceptable cables are Tektronix part
numbers 012-0991-00, -01, -02, and -03. In order to maintain the EMI performance conforming to the above regulations,
the following cables, or their equivalent, should be used: a shielded Centronics cable, 3 meters in length, part number
012-1214-00, and a shielded RS-232 cable, 2.7 meters in length, CA part number 0294-9.