Performance Tests
4–82 TDS 684A, TDS 744A, & TDS 784A Service Manual
HDisconnect the signal source from CH1, wait a few seconds, then
reconnect the signal.
HCONFIRM that the TRIG’D LED is lighted and the waveform is
stable.
HDisconnect all test equipment (TSG 121) from the digitizing
oscilloscope.
HPress save/recall SETUP, the main-menu button Recall Factory
Setup, and the side-menu OK Confirm Factory Init.