Specifications
TDS3000 Series Service Manual 1–5
Table 1–1: Specifications (cont.)
Horizontal
Acquisition (horizon-
tal) resolution Normal Fast trigger
Record length 10,000 points 500 points
Acquisition rate,
maximum Up to 450 waveforms/s Up to 3,000 waveforms/s
S
a
mp
l
e
r
at
e
r
ang
e
Acquisition res-
olution TDS3012
TDS3014 TDS3032
TDS3034 TDS3052
TDS3054
Normal 100 S/s to
1 GS/s 100 S/s to
2.5 GS/s 100 S/s to
5 GS/s
Fast trigger 5 S/s to
1.25 GS/s 5 S/s to
2.5 GS/s 5 S/s to
5 GS/s
Seconds/division
range 4 ns/div to
10 s/div 2 ns/div to
10 s/div 1 ns/div to
10 s/div
Sample rate and
delay time accuracy ±200 ppm over any 1 ms time interval
Trigger
External trigger input,
typical 1 M in parallel with 17 pF, TekProbe compatible
(TDS3012, TDS3032, TDS3052 only)
Ext
er
nal t
r
igg
er
mmm e
Overvoltage category Maximum voltage
m
axi
m
u
m
voltag
e
CAT I Environment (refer to
page 1–11) 150 VRMS (400 Vpk)
CAT II Environment (refer to
page 1–11) 100 VRMS (400 Vpk)
For steady-state sinusoidal waveforms, derate at 20 dB/decade above
200 kHz to 13 Vpk at 3 MHz and above.
External trigger maxi-
mum floating voltage 0 V from chassis (BNC) ground to earth ground
30 VRMS (42 Vpk) only under these conditions: no signal voltages
>30 VRMS (>42 Vpk), all common leads connected to the same voltage,
no grounded peripherals attached
Edg
e
t
r
igg
er
se s
Source Sensitivity
se
n
s
itivity Any channel, DC coupled 0.35 div from DC to 50 MHz,
increasing to 1 div at oscilloscope
bandwidth
Edg
e
t
r
igg
er
se s p
Source Sensitivity
se
n
s
itivity, ty
p
ical External trigger 100 mV from DC to 50 MHz,
increasing to 500 mV at 300 MHz
External/10 trigger 500 mV from DC to 50 MHz,
increasing to 3 V at 300 MHz