Performance Verification
TDS3000 Series Service Manual 4ā5
This test checks the DC voltage measurement accuracy in the average acquisition
mode.
1. Set the DC voltage source output level to 0 V.
2. Connect the DC voltage source to the oscilloscope channel 1 input as shown
below.
DC voltage
source
Ch 1
TDS3000 Series Oscilloscope
3. Push the SAVE/RECALL menu button.
4. Push the Recall Factory Setup screen button and then push the OK
Confirm Factory Init screen button.
5. Push the acquire MENU button.
6. Push the Mode screen button and then push the Average screen button.
7. Adjust the number of averages to 16 with the general purpose knob.
8. Skip to step 11.
9. Move the DC voltage source output cable to the oscilloscope channel you
want to check.
10. Push the channel button (CH 1, CH 2, CH 3, or CH 4) for the channel you
want to check.
11. Push the MEASURE menu button.
12. Push the Select Measurement screen button.
13. Push the āmoreā screen button until you can select the Mean measure-
ment.
14. Push the vertical MENU button.
Check DC VoltageMeasurement Accuracy