TDS3000 Series Service Manual 5–1
Adjustment ProceduresThis chapter contains the factory adjustment procedures for the oscilloscope.
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures. Also refer to the chapter Operating Information for information
about using the oscilloscope.
NOTE. The voltage references inside the oscilloscope are very stable over time
and should not require routine adjustment. Before performing any procedure in
this chapter, first verify that the oscilloscope does not meet specifications. Refer
to the chapter Performance Verification for procedures to verify the specifica-
tions.
Required Equipment
The following equipment, or a suitable equivalent, is required to complete these
procedures.
Description Minimum
requirements Example
DC voltage source 50 mV to 50 V, ±0.1% ampli-
tude accuracy
W
av
e
t
e
k
9500
Os
cillo
s
co
pe
Calibrator with one 9510
Op e
Leveled sine wave generator 30 kHz to 50 MHz, 100 ppm
frequency accuracy, 3% ampli-
tude accuracy
O
ut
p
ut Modul
e
Fast-rise generator 1 MHz repetition rate, 1 V
Pk-Pk signal, ≤1 ns rise time
50 BNC cable
(five required 1)BNC male to BNC male,
≈ 10 in (25 cm) long Tektronix part number
012-0208-00
BNC T (three required 1)One male and two female
BNC connectors Tektronix part number
103-0030-00
BNC female-to-female adapter Two female BNC connectors Tektronix part number
103-0028-00
Adjustment tool Non conducting shaft with
diameter ≤2.5 mm (≤0.1 in) Tektronix part number
003-1433-00
1The number required is reduced by two if you are adjusting a two-channel
oscilloscope.