4 Operations

Model 356WA

 

 

 

 

Heater

Cell Compartment Terminal Strip

Control Assembly

 

 

 

 

 

1

2

3

10

20

30

 

 

 

 

Span

 

Range

 

Humidifier

1

2

3

4

5

 

 

 

Control

 

 

 

 

 

Current Adj

Alarm 1

 

2 Amp

 

Cell

 

 

 

 

 

 

 

 

 

 

Assembly

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Alarm 2

 

 

 

 

Heater

 

 

 

 

 

 

 

 

 

 

Flowmeter

 

 

 

 

 

 

 

 

 

 

Humidifier

 

 

 

 

 

 

 

 

 

 

Column

 

 

 

 

 

 

 

 

 

 

Assembly

 

 

 

 

 

 

 

 

 

 

Sample

 

 

 

 

 

 

 

 

 

 

System

 

 

 

 

 

 

 

 

 

 

Assembly

 

 

 

 

 

 

 

 

 

 

Figure 4-1: Typical Model 306WA With 2 Alarm Option

Interconnect

Terminal Strip

Throttle

Valve

Sample System Assembly with Reservoir

WARNING: Type A electrolyte is caustic. Use extreme care in handling. Protective equipment including but not limited to gloves and safety glasses should be worn while handling electrolyte. Refer to the Material Safety Data Sheet in the Appendix regarding potential hazards and corrective action in case of accident.

Type B electrolyte is a 20 % solution (w/v) of potassium carbonate and should be used when the CO2 level in the background gas is between 500 ppm and 1 %. This narrow range is rarely encountered. It is, however available from the factory. Safety related information for this electrolyte can be found in the Appendix.

Type C electrolyte is a 20 % solution (w/v) of potassium bicarbonate and should be used when the CO2 level in the background gas is between 1 and 100 %. Safety related information for this electrolyte can be found in the Appendix.