4-1
TT
TT
Trace Oxygrace Oxyg
race Oxygrace Oxyg
race Oxygen Analen Anal
en Analen Anal
en Analyzyz
yzyz
yzerer
erer
er Operations 4Operations 4
Operations 4Operations 4
Operations 4
Teledyne Analytical Instruments
OperationsOperations
OperationsOperations
Operations
4.14.1
4.14.1
4.1 Filling the ReservoirFilling the Reservoir
Filling the ReservoirFilling the Reservoir
Filling the Reservoir
The reservoir is located on the right side of the analyzer case.
1) Insure that the cap on the drain spout is securely tightened.
2) Remove cap from fill port on top of reservoir.
3) Pour distilled water into reservoir until it is half full (about one
quart). The water will automatically flow into the humidifier
column.
4) Replace cap on fill port and securely tighten. A missing or loose
cap will permit the sample to vent into the analyzer case.
4.24.2
4.24.2
4.2 Detector CellDetector Cell
Detector CellDetector Cell
Detector Cell
The cell is located in the heated compartment on the left side of the
analyzer case, as shown in Figure 4-1. To open the compartment, unscrew
the captive knurled knobs at the top and bottom of the compartment and
remove the plastic window.
4.2.14.2.1
4.2.14.2.1
4.2.1 Cell PackagingCell Packaging
Cell PackagingCell Packaging
Cell Packaging
The cell is packaged separately from the analyzer. It is filled with
distilled water to prevent oxidation of the electrodes from exposure to the
atmosphere. The cell should be left filled with the distilled water until the
analyzer is installed and ready for operation. The cell should not be exposed
to the atmosphere for any prolonged duration.
4.2.24.2.2
4.2.24.2.2
4.2.2 ElectrolyteElectrolyte
ElectrolyteElectrolyte
Electrolyte
The cell electrolyte is Teledyne Type A, used in applications where
there is a complete absence of acidic anhydrides (CO2, SO2) in the sample
gas. Type A electrolyte is a 10% solution (w/v) of reagent-grade potassium
hydroxide (KOH) in distilled water.