Trace Oxygen Analyzer

Appendix

 

 

 

 

From these computations, the exit flow rate should be set to 189 cc/min. using an appropriate flow measuring device and the target respositioned if necessary so that the float is centered within the target opening.

Subsequently, it should not be necessary to measure the exit flow unless another circumstance, of the type listed above, occurs.

Teledyne Analytical Instruments

A-7

Page 41
Image 41
Teledyne 356WA instruction manual Trace Oxygen Analyzer Appendix