Trace Oxygen Analyzer

Appendix

 

 

 

 

Appendix

Specifications

Standard Ranges:

Three ranges between 0-10 PPM and

 

0-5000 PPM O2 (0-2 PPM O2 avail-

 

able as an option).

Sampling System:

Wetted parts: 304 welded stainless

 

steel.

Sensitivity:

1% of low range.

Accuracy:

+2% of low range.

Response and Recovery Time:

90% in less than 1 minute (for lowest

 

range).

Operating Temperature Range:

+40 °F to +120 °F (+5 °C to +49 °C).

Sample Requirement:

Flow: 150 cc/min.*

 

Pressure: 1 to 150 psig.

 

Temperature: +60 °F to +100 °F

 

(+15 °C to +38 °C)

Power Requirement:

115 VAC, 50/60 Hz, 150 W (other

 

voltages available.)

Alarms:

Adjustable single or dual alarm (op-

 

tional) setpoints.

Recorder Signal Output:

Voltage: 0-1 VDC or less.

 

Current: 1-5, 4-20, 10-50 mADC

 

(optional.)

Local Readout:

Digital or analog meter.

*Specified flow rate required only during calibration. Measuring cell is not sensitive to changes in flow rate.

Teledyne Analytical Instruments

A-1

Page 35
Image 35
Teledyne 356WA instruction manual Trace Oxygen Analyzer Appendix Specifications