Trace Oxygen Analyzer

Operational Theory 2

 

 

 

 

Operational Theory

2.1 Sensor

The sensor is an open-cathode cell, an electrochemical transducer specific to oxygen. The cathode of the cell is composed of silver screen elements with a large surface area. The screen assembly is mounted in an acrylic block, with the lower edges of the screens immersed in potassium hydroxide electrolyte. A thin layer of electrolyte is maintained on the sur- faces of the screens by capillary action. A lead disk is positioned under the screens and serves as the anode. An exploded view of the cell is showm in Figure 5-1.

The sample gas stream is passed directly over the cathode screens, initiating an electrochemical reaction. Four electrons are generated by the oxidation of the lead anode, and are then used to reduce oxygen at the cathode. The flow of electrons between the anode and cathode creates an electric current which is directly proportional to the oxygen concentration in the sample stream. In the absence of oxygen, no oxidation or reduction takes place, and no current is produced.

In simplified form, the reaction may be described as follows: oxygen is reduced at the cathode by the mechanism

4e- +O2 + 2H2O 4OH-

This cathodic half-reaction occurs simultaneously with the anodic half- reaction

Pb + 2OH- PbO + H2O + 2e-

The overall reaction is

O2+ 2Pb 2PbO

2.2 Humidifier

It is necessary to maintain a film of electrolyte on the screens of the electrode assembly. This means that the humidity of the sample as it flows through the cell must be such that the water vapor pressure of the electro-

Teledyne Analytical Instruments

2-1

Page 7
Image 7
Teledyne 356WA instruction manual Sensor, Humidifier