Operation

4.4.1 Default Parameters

The versatility of this analyzer usually results in significant changes being made to parameters over the course of time to better suit a particular application. Occasionally processes change requiring alteration to alarms, ranges etc. At some time, it may be beneficial to reset the analyzer to the default conditions as it was when shipped from the factory. Below is a listing of the default parameters used in configuring your instrument:

Range/Application: Refer to the data sheet on the first page of this manual

Range: Auto

Alarms: Enabled, AL1=20ppm, AL2=50ppm, HI,

NON-FAILSAFE, NON-LATCHING,

THC levels

Auto Span Timing: Not available.

THC Span: 8 ppm methane in air.

Password: TAI

Analog Outputs: Ch1=THC levels

Ch2=Range ID

4.4.2 Style Conventions

The following typeface conventions are used when referring to screen names, key presses and screen readout:

Screens:

Arial 12 pt. type in capital letters.

 

Example: ANALYZE or MAIN

 

screen or menu.

Key presses:

<key> The particular keystroke to

 

enter is placed between < and >.

 

Example: <enter> or <escape> or

 

<▲> (UP key) or <▼> (DOWN

 

key.

 

Only when the keystroke is to be

 

entered will it be placed between

 

the brackets. If discussing a

Teledyne Analytical Instruments

31

Page 33
Image 33
Teledyne SERIES 4030, HEATED THC manual Default Parameters, Style Conventions