Operation
________________________________________________
<Enter or Escape> Start the diagnostic testing routine by pressing <Enter> or cancel out by pressing <Escape>.
If you pressed <Enter> the
To return the analyzer to the MAIN MENU, press <Enter> after the results screen.
If you pressed <Escape> you will be returned to the Analyze Mode.
4.4.10 Span
The SPAN function is used to calibrate the analyzer.
The analyzer is calibrated using span gas as described in Section 4.1. This section assumes that this gas has been properly connected and the line checked for leaks.
To initiate a pan calibration:
<▲▼> From the MAIN MENU, scroll down to the SPAN function.
<Enter> Press <Enter> to activate the SPAN function.
________________________________________________
Span? | YES | ||
SMP | Span? | NO | |
THC | Sp: | 8.00 | ppm |
SMP | Sp: | 0.00 | ppm |
Span Begin
________________________________________________
Teledyne Analytical Instruments | 46 |