Operation

11.MODIFIABLE ITEM screens have a flashing value somewhere in the line.

4.4.4.1 MAIN MENUS

The MAIN MENU consists of 13 functions you can use to customize and check the operation of the analyzer. They are listed here with brief descriptions:

1.AUTO-CAL:Used to define and/or start an automatic calibration sequence. AUTOCAL is not available for standard configurations of the 4030.

2.GAIN: Used to change the amplification of the incoming signal. This is factory set, and should not change.

3.PASSWORD: Used to establish password protection or change the existing password.

4.LOGOUT: Logging out prevents unauthorized tampering with the analyzer settings.

5.MODEL: Displays Manufacturer, Model, and Software version of the instrument.

6.SELF-TEST:The instrument performs a self-diagnostic routine to check the integrity of the power supply, and output boards.

7.SPAN: Set up and/or start a span calibration

8.ALARMS: Used to set the alarm set points and determine whether each alarm will be active or defeated, HI or LO acting, and failsafe or not.

9.RANGE: Used to set up three analysis ranges that can be switched automatically. With auto-ranging or used as individual fixed ranges.

10.ANALOG-OUT ADJUST: Adjust the analog output with offset and scaling for Channel 1 only.

11.CHANGE STREAM: Not available.

12.TIMING: Low-level application specific timing parameters. Factory set. Do not change these.

13.STANDBY: Removes power to outputs and displays, but maintains power to internal circuitry.

Any function can be selected at anytime. Just scroll through the MAIN MENU with the DOWN/UP keys to the appropriate function, and

Teledyne Analytical Instruments