Teledyne HPM-2002-OBE Device Status, Default RS232/485 Specifications, 9600, Character Length

Models: HPM-2002-OBE

1 39
Download 39 pages 52.25 Kb
Page 20
Image 20

Restore Factory Defaults

/#<CR>

Restore Calibration Register Default Values

Set Zero

/0 <CR>

Store Present Pressure as Instrument ZERO

 

 

(if Piezo <32 Torr & Pirani <50 mTorr)

Force Zero

/!<CR> Override Limit Checks and Store Present

 

 

Pressure as Instrument ZERO

3.3.7.Device Status

When requested to transmit its status the HPM-2002-OBE responds with a five digit number which is explained in the following:

digit#:

1

2

3

4

5

Serial Receiver Overflow

 

 

 

 

 

 

 

 

 

 

=4

 

 

 

 

 

 

 

 

Main Board EEPROM Error

=2

 

 

 

 

 

 

 

 

Probe EEPROM (any) Error

=1

 

 

 

 

 

 

 

 

Probe EEPROM Not Responding

 

 

 

 

 

 

 

 

 

 

 

 

=8

 

 

 

 

 

 

Probe EEPROM Read Error

 

 

=1

 

 

 

 

 

 

Probe EEPROM Checksum Error

 

 

 

 

 

 

 

 

 

 

 

 

 

 

=4

 

 

 

 

Probe EEPROM Verification Error

 

 

 

 

=2

 

 

 

 

Probe EEPROM Identification Error

 

 

 

 

=1

 

 

 

 

Communications Syntax Error

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

=4

 

 

Piezo Sensor Bad (voltage out of range)

 

 

 

 

 

 

=2

 

 

Pirani Sensor Bad (voltage out of range)

 

 

 

 

 

 

=1

 

 

High Setpoint Alarm (pressure exceeds setpoint)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

=4

Low Setpoint Alarm (pressure less than setpoint)

 

 

 

 

 

 

=2

Gas# Changed (not the same as when unit last calibrated)

 

 

 

 

=1

3.3.8.Default RS232/485 Specifications

Baud Rate

9600

Character Length

Eight data bits

Parity

None

Stop Bits

1

RS 485 Universal Address

00

HPM-2002-OBEVacuum Gauge

Page 20 of 39

Page 20
Image 20
Teledyne HPM-2002-OBE Device Status, Default RS232/485 Specifications, 9600, Character Length, Eight data bits, None