Restore Factory Defaults  | /#<CR>  | Restore Calibration Register Default Values  | 
Set Zero  | /0 <CR>  | Store Present Pressure as Instrument ZERO  | 
  | 
  | (if Piezo <32 Torr & Pirani <50 mTorr)  | 
Force Zero  | /!<CR> Override Limit Checks and Store Present  | |
  | 
  | Pressure as Instrument ZERO  | 
3.3.7.Device Status
When requested to transmit its status the 
digit#: | 1  | 2  | 3  | 4  | 5  | |||||
Serial Receiver Overflow  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
=4  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | ||
Main Board EEPROM Error  | =2  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | |
Probe EEPROM (any) Error  | =1  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | |
Probe EEPROM Not Responding  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | =8  | 
  | 
  | 
  | 
  | 
  | 
  | ||
Probe EEPROM Read Error  | 
  | 
  | =1  | 
  | 
  | 
  | 
  | 
  | 
  | |
Probe EEPROM Checksum Error  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | =4  | 
  | 
  | 
  | 
  | ||
Probe EEPROM Verification Error  | 
  | 
  | 
  | 
  | =2  | 
  | 
  | 
  | 
  | |
Probe EEPROM Identification Error  | 
  | 
  | 
  | 
  | =1  | 
  | 
  | 
  | 
  | |
Communications Syntax Error  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | =4  | 
  | 
  | ||
Piezo Sensor Bad (voltage out of range)  | 
  | 
  | 
  | 
  | 
  | 
  | =2  | 
  | 
  | |
Pirani Sensor Bad (voltage out of range)  | 
  | 
  | 
  | 
  | 
  | 
  | =1  | 
  | 
  | |
High Setpoint Alarm (pressure exceeds setpoint)  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | 
  | ||
  | 
  | 
  | 
  | 
  | 
  | =4  | ||||
Low Setpoint Alarm (pressure less than setpoint)  | 
  | 
  | 
  | 
  | 
  | 
  | =2  | |||
Gas# Changed (not the same as when unit last calibrated)  | 
  | 
  | 
  | 
  | =1  | |||||
3.3.8.Default RS232/485 Specifications
Baud Rate  | 9600  | 
Character Length  | Eight data bits  | 
Parity  | None  | 
Stop Bits  | 1  | 
RS 485 Universal Address  | 00  | 
 | Page 20 of 39  |