1.5. HPM-2002-OBE RS232/485 Output Module

The RS232/485 module consists of a power conversion/sensor transducer board, microprocessor board and a user interface option board. Serial communication is conducted via an RJ-11 type connector. With RS232 option (EIA-232 Rev. E) communication can easily be established between the module and the serial port of a PC. The RS485 option allows the user to address multiple units and allows operation at distances of up to 4000 feet. The RS232/485 module includes a four digit floating-point LED display.

1.6. HPM-2002-OBE DeviceNet Module

The DeviceNet module consists of a power conversion/sensor transducer board, microprocessor board and a user interface option board. All communication is conducted via a 5-pin “Micro” style with center pin, male pin contacts. The module has passed the ODVA DeviceNet test and conforms to the vacuum/pressure gauge device profile, an Electronic Data Sheet (EDS) with limited features is also available upon request.

1.7. Specifications

Measuring range

1x10-4to 10+3 Torr

..............................................................................................................................

1.3x10-4to 1.3x10+3 mbar

Accuracy (N2, T=23°C)

+ 20% of reading (1 x10-3to 50 Torr)

............................................................................................................

+ 1.5% of reading (50 to 1000 Torr)

Ambient temperature operating range

0° to 50°C

Process control

2 TTL outputs (1 TTL remote zero command input)

Digital readout

Four digit LED

Equipment operating ranges

12-30 VDC

Transducer mounting

Any position without recalibration.

Transducer internal volume

< 1.5 cc

Wetted material

Au, Si3N4, Si, PyrexTM,

....................................................................................................................

KovarTM, 316 stainless steel and

....................................................................................................

High Temp/Low Outgassing UHV Epoxy

Weight (OBE & HPM-2002s tube)

12 oz (1/8” NPT tube)

Calibrated for nitrogen

Conversion Factors for other gases are selectable

Burst Pressure (Tube)

150 psig

Proof Pressure (*)

30 psig

Nominal Operating Pressure (Tube)

1x10-4to 10+3 Torr

Input Operating Range

11.5 to 30 VDC

Input Power

24 VDC @ 125 mA

* Maximum pressure above which may cause permanent damage.

HPM-2002-OBEVacuum Gauge

Page 6 of 39

Page 6
Image 6