2. Installation
This section is designed to assist in getting a new pressure gauge into operation as quickly and easily as possible. Please read the following instructions thoroughly before installing the instrument.
2.1. Receiving Inspection
Carefully unpack the Hastings Model
transducer tube (part #HPM
Compare each component shipped against the packing list. Ensure that all parts are present (i.e. transducer, electronics module, hardware, etc.). Optional equipment or accessories will be listed separately on the packing list.
2.2.Quick Start
•Unpack and inspect all items for any obvious signs of damage due to shipment. Immediately advise the carrier who delivered the shipment if any damage is suspected.
•Wire the
•Using a unipolar DC Power Supply, set the desired operating voltage within the range of 12 VDC to 30 VDC.
•Connect the
•Note that the connector is keyed.
•A finger tight connection is all that is required for adequate operation.
•Transducer tube may be installed in any orientation. However, if condensation is likely to occur, then the tube port should be orientated downward.
•When installing 1/8” NPT style transducer tube, use the 7/16” wrench flats.
•Attach cable.
•With the vacuum chamber at atmosphere, turn on the power supply (typically +24 VDC. Gauge is now reading pressure.
•For best accuracy, the gauge should now be zeroed. Pump the vacuum system down to low
•To set the “Zero”, place the
•Using a small flat head screwdriver, rotate the “ADJUST” rotary encoder until the unit flashes between “0.0” and
•Return to the “RUN” mode by using the “SELECT” button
2.3.Transducer Installation
The transducer tube may be installed in any orientation. Although the transducer tube is rugged and will perform well in many harsh environments, the tube should be installed in a clean and careful manner. The tube is configured with the vacuum fitting requested. If your vacuum environment is highly contaminated or has unique fitting requirements, a Hastings filter or special adapter may be needed. Please contact the Hastings Instruments Sales Department for assistance in your system configuration.
2.4.OBE Module Installation
•Environment:
•Indoor use
•Altitude up to 2000 meters
•Operating temperature range from 5 to 40°C
Page 7 of 39 |