Teledyne HPM-2002-OBE instruction manual Initial Operation, Volt, 4-20 mA, RS232/485

Models: HPM-2002-OBE

1 39
Download 39 pages 52.25 Kb
Page 8
Image 8
0-10 Volt

Maximum relative humidity: 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C.

Installation category II

HPM-2002-OBE

CABLE PINOUT

 

Analog Ouput

Analog Ouput

Digital Output

Pin

0-10 Volt

4-20 mA

RS232/485

No.

 

 

 

 

 

 

 

1

High Setpoint Output

High Setpoint Output

High Setpoint Output

2

Low Setpoint Output

Low Setpoint Output

Low Setpoint Output

3

Power Input (+12 to 30 VDC)

Power Input (+12 to 30 VDC)

Power Input (+12 to 30 VDC)

4

Power Return

Power Return

Power Return

5

Channel 1 Vout (+)

Channel 1 Iout (-)

NC

6

Analog Return

Analog Shield

NC

7

Channel 2 Vout (+)

Channel 2 Iout (-)

NC

8

Digital Ground

Digital Ground

Digital Ground

9

Remote Zero

Remote Zero

Remote Zero

10

NC

Channel 1 Iout (+)

NC

11

NC

Channel 2 Iout (+)

NC

12

NC

NC

NC

13

NC

NC

NC

14

Rx In (Internal Use)

Rx In (Internal Use)

NC

15

Tx Out (Internal Use)

Tx Out (Internal Use)

NC

2.5. Initial Operation

Upon applying power to the control unit a pressure measurement will be given in Torr for nitrogen. However, it is recommended that the user follow the instructions for zeroing and adjusting the output at atmosphere.

HPM-2002-OBEVacuum Gauge

Page 8 of 39

Page 8
Image 8
Teledyne HPM-2002-OBE instruction manual Initial Operation, Volt, 4-20 mA, RS232/485