EAR & FOREHEAD THERMOMETER 201
12 FEATURES
Model | |
Temperature Sensor type | Infrared |
Measuring temperature range | 32 °C ~ 42,9 °C (89,6 °F ~ 109,3 °F) |
Room temperature | 5 °C ~ 59,9°C (41 °C ~ 139,8 °F) |
Display accuracy | 0,2°C or 0,4°F |
Measurement accuracy | 35,5°C ~ 42°C - (+/- 0.2°C) |
| 32° C to 35.4 °C - (+/- 0.3 °C) |
| 42,1C to 42.9 C° - (+/- 0.3 °C) |
| 95,9°F ~ 107,6°F |
| 89,6°F ~ 95,8°F - (+/- 0,5°F) |
| 107,7°F ~ 109,3°F - (+/- 0,5°F) |
| Room temperature +/- 1°C |
Operating ambient temperature range 15°C ~ 40°C (59,0°F ~ 104,0°F)
Operating ambient humidity range | 30 ~ 80% RH | |||
Storage environment temperature | ||||
range |
|
| ||
Storage environment humidity range | 30 ~ 85% RH | |||
Memory | Last 30 measurements | |||
Battery Type | 1 x CR2032 - DC 3V - Lithium | |||
Automatic | 30 seconds after last measurement | |||
Weight | 65g (with battery) | |||
Classification |
|
| • | Type BF |
|
|
| • | Class IIa |
|
|
| • | Internally powered equip- |
|
|
|
| ment |
|
|
| • | IPX0 |
|
|
| • | Not suitable for use in pres- |
|
|
|
| ence of flammable anes- |
|
|
|
| thetic mixture with air or |
|
|
|
| with oxygen or nitrous |
|
|
|
| oxide |
|
|
| • | Continuous operation with |
|
|
|
|
10