Display Handling:

Display mounting plate lists theYORK part numbers of the Display supporting components mounted on the Display mounting plate and the required Microboard Program Jumper (JP2 through 8) configurations. Microboard Program Jumpers

JP2 - JP8 will have to be configured appropriately for the replacement display. Refer to Table 2 “Program Jumpers”.

Display Handling:

1.The display is made of glass. It could break if dropped.

2.The display front surface is easily scratched. If soiled, wipe with a dry cotton cloth. Use no water or chemicals.

3.The display is static sensitive. Electrostatic dis- charges may damage the display.

4.A laminated film is adhered to the display front glass surface to prevent it from being scratched. Peel off very slowly to prevent static damage.

Always remove control power from the OptiView Remote Control Center before connecting or disconnecting wires to the display. Connecting or disconnecting wires to the display with power applied will damage the display!!!

BACKLIGHT LAMP REPLACEMENT:

SHARP LQ10D367 Display: (Refer to Fig. 41) Removal:

The Lamp slides into the Display from left to right and is secured with a locking tab.

1.Remove Control Power from the OptiView Control Center.

2.Remove protective cover from rear of Display.

3.Disconnect Lamp AC power connector from Backlight Inverter Board.

FORM 50.40-OM2

4.Using fingernail or thin flat blade screwdriver, bend the locking tab outward slightly to clear the Lamp housing protrusion.

5.Grasp Lamp AC power connector and gently pull until Lamp housing clears locking tab.

6.Grasp Lamp housing and pull until Lamp housing is completely removed from the Display.

Installation:

1.Slide new Lamp into Display from left to right until Lamp housing protrusion locks into Display locking tab.

2.Connect Lamp AC power connector to Backlight Inverter Board.

3.Apply Control Power to OptiView Control Center.

LG Semicon LP104V2 Display (refer to Fig. 42) Removal:

The Lamp slides into the Display from left to right and is secured with a screw.

1.Remove Control Power from the OptiView Remote Control Center.

2.Remove protective cover from rear of Display.

3.Disconnect Lamp AC power connector from Back- light Inverter Board.

4.Using small Phillips screwdriver, remove lamp retaining screw.

5.Grasp Lamp AC power connector and gently pull until Lamp housing is completely removed from the Display.

Installation:

1.Slide new Lamp into Display from left to right until Lamp housing is fully inserted.

2.Secure Lamp with Lamp retaining screw.

3.Connect Lamp AC power connector to Backlight

Inverter Board.

4. Apply AC power to OptiView Remote Control 5 Center.

YORK INTERNATIONAL