VXI EX7000 user manual Trigger Bus Connectors, Digital I/O Connector

Page 20

VXI Technology, Inc.

Trigger Bus Connectors

The EX7000 provides an LXI compatible trigger bus connector set on the rear panel of the instrument. Both connectors have the same pin assignments and are wired in parallel. For more information on the LXI Trigger Bus and creating an appropriate cable, please visit www.lxistandard.org and refer to LXI Standard Revision 1.2 and LXI Trigger Bus Cable and Terminator Specifications Rev 1.1. The trigger bus pin assignments are provided on the next page.

Trigger Bus Connectors

Pin

Signal

Pin

Signal

1

+3.3 V

14

RP_TRIG_P0

2

GND

15

RP_TRIG_N0

3

RP_TRIG_P1

16

RESERVED

4

RP_TRIG_N1

17

RP_TRIG_P2

5

GND

18

RP_TRIG_N2

6

RP_TRIG_P3

19

GND

7

RP_TRIG_N3

20

RP_TRIG_P4

8

GND

21

RP_TRIG_N4

9

RP_TRIG_P5

22

GND

10

RP_TRIG_N5

23

RP_TRIG_P6

11

RESERVED

24

RP_TRIG_N6

12

RP_TRIG_P7

25

RESERVED

13RP_TRIG_N7

1

13

14

25

Digital I/O Connector

The EX7000 features an 8-channel digital I/O port on the rear panel of the instrument. This port can be used as a trigger source and as a general purpose output device. The digital I/O connector is a standard DB-9 with the following pin assignment:

Digital IO Connector

Pin

Signal

Pin

Signal

1

DIO Channel 0

2

DIO Channel 1

3

DIO Channel 2

4

DIO Channel 3

5

DIO Channel 4

6

DIO Channel 5

7

DIO Channel 6

8

DIO Channel 7

9GND

1

5

6

9

Mating Connector: Tyco 5-747904-2

As a general purpose output device, each DIO channel can be independently programmed with regards to its output functionality and its static level to assume when enabled as an output. When not enabled as an output, a channel becomes tri-stated, preventing conflict with other potential voltage drivers. Reference the port’s electrical specifications in Table 3-1 for voltage tolerance limits and output drive capabilities. Regardless of output functionality, each channel provides constant input functionality. That is, the input level on each channel can be accessed without a specific enable command. Moreover, the base functionality of the DIO channels is not affected by triggering, scanning, or any other instrument process.

20

EX7000 Series Basic Operation

Image 20
Contents EX7000 Series VXI Technology, Inc EX7000 Series Preface Table of Contents Section IndexWarranty Limitation of WarrantyCertification Restricted Rights LegendEMC Steve Mauga, QA ManagerService should only be performed by qualified personnel Terms and SymbolsSupport Resources Section IntroductionCommon Features EX7400 LXI Based Custom RF Interface PlatformInteractive Control Populated EX7400 MainframeRelay Drive Control 4JAVA-BASED Graphical User InterfaceConfiguration Table Relay OdometerExclude Lists Scan ListsEX7000 Specifications EX72SFUnpacking Installation LocationOverview Network Configuration Driver InstallationEX7000 Hardware Setup MAC Address AutoIP Default Address VXI Technology, Inc EX7000 Series Preparation for Use Rear Panel View EX7000 Typical Rear Panel View EX7400Trigger Bus Connectors Digital I/O ConnectorCharacteristic Value General Purpose I/OReset Button Front Panel ViewEX72SF EX7400 WEB page Operation General WEB page OperationPassword Protection EX7000 Main WEBVXI Technology Logo EX7000 Title BlockSoft Front Panel EX7000 Soft Front PanelA pseudorandom fashion until it finds one that is available Dhcp or static IP configurations Time Configuration EX7000 Time Configuration WEB page ManualTime Configuration WEB page Sntp Upgrade Device IdentifyReset RebootTheory of Operation System OverviewRelay Drive Connector Banks OPEN-COLLECTOR VS. TTL Drive LinesRelay Access & Control Continuous ModePulse Mode Continuous Operation ModeIndex