VXI Technology, Inc.
RDBs can either be commanded one board at a time or several RDBs may be commanded simultaneously. The RDB has 72 relay control drive lines capable of sinking 200 mA and twelve very
•six
•six indicator/status bits
•one very
•four relay ID bits
•one relay power input
•one 5 V input
•one ground
NOTE All devices connected to a particular bank (i.e. group of four connectors) must be of the same type: continuous or latching. Mixing of different type may result in erratic behavior and damage to overall system.
OPEN-COLLECTOR VS. TTL DRIVE LINES
The EX7300 and EX7400 relay drive connectors reside on a (RDBs) that are partitioned into three banks: J2 through J5 constitute bank 0, J6 through J10 bank 1, and J11 through J14 bank 2. Although a single RDB may contain both latching and
Relay Driver Board (RDB)
J13 J11
J12 J10
J9 J7
J8 J6
J5 J3
J4 J2
Some connectors have been removed for clarity. | Bank 2 | Bank 1 | Bank 0 |
FIGURE 5-2: RELAY DRIVE CONNECTOR BANKS
Additionally, the RDB of the EX7300 and EX7400 may have varying quantities of
36 | EX7000 Series Web Page Operation |