2.2 Jumper Setting

There is only one jumper on the OME-PCI-1002. JP1 is used to select the analog input type. For single-ended inputs, users should connected pin 1, 3 and pin 2, 4. For differential inputs, Pin 3, 5 and Pin 4, 6 should be connected.

JP1JP1

1

2

1

 

 

5

6

5

 

Single-ended

Differential

Inputs (Default)

Inputs

2

6

2.3 A/D Calibration

• A/D Calibration for OME-PCI-1002 card

 

 

 

 

 

 

 

Step 1: Apply +10V to channel 0.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Step 2: Apply +0V to channel 1.

 

 

 

 

 

 

 

Step 3: Apply -10V to channel 2.

 

 

 

 

 

 

 

Step 4: Run DEMO6.EXE.

 

 

 

 

 

 

 

Step 5: Adjust VR1 until channel 0 = fff or ffe

 

 

 

 

 

 

 

VR1, VR2, VR3

Step 6: Adjust VR2 until channel 1 = 800 or 801

 

 

 

 

 

 

 

Step 7: Adjust VR3 until channel 2 = 000 or 001

 

 

 

 

 

 

 

Step 8: Repeat Step 4 & Step 5 & Step 6 until all read

 

 

 

 

 

 

 

properly.

 

 

 

 

 

 

 

14

Image 14
Contents User’sGuide Servicing North America Table of Contents Software and Demo Program Diagnostic Program Introduction General DescriptionBlock Diagram X86 SystemFeatures Analog Inputs SpecificationsPower Consumption 3 D/I and D/O 4 A/D Trigger MethodsProgrammable Timer/Counter Interrupt ChannelApplications OME-PCI-1002 series multifunction cardsProduct Check List Release NotesHardware Configuration Board LayoutJumper Setting A/D CalibrationSystem Block Diagram OME-PCI-1002 System Function BlockDaughter Boards OME-DB-8225 OME-DB-37OME-DB-16P Isolated Input Board OME-DB-16R Relay Board Analog Input Signal Connection Page Page Signal Shielding Jumper Settings Connectors Page How to Find the I/O Address O RegistersPage I/O Address Map Section 8254 Registers Status Register Clear Interrupt General Control Register Interrupt Source SelectionTrigger Method Selection Function Operation Digital I/O8254 Timer A/D triggerSymbol Name Minimum Maximum A/D Conversion CPUSuggested Settling Times 1 A/D Conversion Trigger Modes Software Trigger and Polling Techniques Page Page Software and Demo Program Power-ON Plug & Play Test Driver Plug & Play TestD I/O Test A/D TestWARRANTY/DISCLAIMER Temperature