Omega 1002 manual Suggested Settling Times

Page 39

Before performing A/D conversions, the JP1 jumper should be set for single-ended or differential.

The software driver supports both polling and interrupt driven A/D, however, polling is the simplest way to perform an A/D conversion.

The settling time of multiplexer depends on the impedance of the signal source. Because the driver does not incorporate the settling time, the user needs to program a delay when switching from one channel to another or when changing gains.

Suggested Settling Times

One settling time is suggested for all ranges of the OME-PCI-1002L. The settling time for the OME-PCI-1002H is based on the analog input range. The table below shows the suggested settling times for each range.

OME-PCI-1002L Settling Time: 33 ∝seconds for all ranges

OME-PCI-1002H Settling Time

Input Range

Settling Time

± 10V

23 ∝S

± 5V

28 ∝S

± 0.1V

140 ∝S

± 0.01V

1300 ∝S

The software driver provides a machine independent timer, P1002_Delay(), for settling time delay. If the user calls P1002_Delay(), counter 0 will be reserved and can not be used as a user programmable timer/counter.

The A/D converter requires a trigger signal to start an A/D conversion cycle. The OME-PCI-1002 supports three trigger modes, software, pacer and external trigger.

39

Image 39
Contents User’sGuide Servicing North America Table of Contents Software and Demo Program Diagnostic Program General Description IntroductionX86 System Block DiagramFeatures Specifications Power ConsumptionAnalog Inputs 4 A/D Trigger Methods 3 D/I and D/OInterrupt Channel Programmable Timer/CounterOME-PCI-1002 series multifunction cards ApplicationsRelease Notes Product Check ListBoard Layout Hardware ConfigurationA/D Calibration Jumper SettingOME-PCI-1002 System Function Block System Block DiagramOME-DB-8225 OME-DB-37 Daughter BoardsOME-DB-16P Isolated Input Board OME-DB-16R Relay Board Analog Input Signal Connection Page Page Signal Shielding Jumper Settings Connectors Page O Registers How to Find the I/O AddressPage I/O Address Map Section 8254 Registers Status Register Clear Interrupt Interrupt Source Selection General Control RegisterTrigger Method Selection Digital I/O Function OperationA/D trigger 8254 TimerSymbol Name Minimum Maximum CPU A/D ConversionSuggested Settling Times 1 A/D Conversion Trigger Modes Software Trigger and Polling Techniques Page Page Software and Demo Program Driver Plug & Play Test Power-ON Plug & Play TestA/D Test D I/O TestWARRANTY/DISCLAIMER Temperature