3. Next, find m & b:

X1

=

0 VDC

ç minimum expected input

=

0 A/D counts

at Transducer minimum output

X2

=

2.0000 VDC

ç maximum allowable input

=

20000 A/D counts

at Transducer maximum output

Y1

=

20.00 %RH

ç choose decimal point = 2

=2000 display counts

Y2

=

90.00 %RH

ç keep decimal point = 2

=

9000 display counts

[Desired % RH resolution 2 places to right

m

=

(Y1 - Y2) / (X1 - X2)

of the decimal point (.01% RH)]

=(2000 - 9000) / (0 - 20,000)

=0.35000

b = Y2 - (m * X2)

=9000 - (0.350 * 20,000)

=002000

4.Programming for the previous example:

 

PRESS 1

 

PRESS V

 

PRESS YES

 

PRESS 035000

 

PRESS YES

 

PRESS 002000

 

PRESS 58

 

PRESS 2

 

THE FINAL CHANNEL

PRESS THE EXIT KEY AND IT IS ECHOED

DESCRIPTION

 

Figure 4-10

OM-550

u36u

Image 41
Contents Table of Contents Date Time ∙ The OM-550 Datalogger may be referred to as Logger IntroductionProduct Usage Warnings Operator Safety InformationFor protection of the OM-550 itself, observe the following Unpacking OM-550 features FeaturesOM-550 Features Activity LED & Horn Keypad ∙ ON-OFF ∙ RUN ∙ OK ∙ Stop∙ PROG/EXIT Application Channel Connectors∙ Optional Connectors To remove the Viewing Window Printer WindowThermal Paper Battery CoverDC Jack Serial Data PortTilt Stand Gain AdjustPower Supplies Voltage Transducer Inputs InputsProbes Probe SETType Thermocouple InputsData Cache Programmed FeaturesStatistics Battery Check POWER-SAVE ModeLOW Battery Operation Battery Installation Power ConnectionConnecting the DC Power Cable Connecting the AC Power AdapterPaper Installation PIN Adapter Data Port Connection∙ Programming outline 1. Select Channel ProgrammingProgramming Keys Position Print FormattingVoltage Channels Temperature ChannelsMath PSEUDO-CHANNEL Not Printing LOG IntervalThis Interval Overrun can be avoided in two ways Operation AT Full Cache Recording Time Rejection Frequency Temperature UnitsTo return to Normal mode Calibration ModeTadjust Matching Transducers to the OM-550 Voltage Input ScalingMatching a Voltage Transducer R1 = 2 R2 = V-2R1 = 2 Matching a Current Transducer= 2.0000 VDC = R1 Data Scaling mX + b calculation Matching and Scaling a Voltage Transducer VDC Example Matching and Scaling a Current TransducerU38u Unit Labels Programmed Information OperationLog Interval Time Temperature Units Enter a Tadjust value, or press OK to finish programming U44u ∙ Performing a Manual Reset Appendix a Calibration ∙ Manual ResetManual Reset programs the OM-550 as follows Setup ∙ Calibration Procedure∙ Thermocouple Adjustment Procedure ∙ Voltage Adjustment ProcedureTo return to normal operation ∙ Exiting CAL ModeAccuracy Appendix B SpecificationsVoltage Measurement Inputs Paper Life PrinterDigital Interface RS-232 ConversionBattery Life POWER/CURRENT RequirementsPhysical Dimensions ConnectorsOM-550 DOS Software Table of Contents Introduction How to Use This ManualFonts Used in this Manual Printer CPU, RAM, Hardware∙ Hard Drive Installation Software InstallationTo start the program Starting the ProgramMaster Menu Reading Data from the OM-550 To load a LOG file Load Existing LOG FilesModify Software Configuration U6u To list data List Data on MonitorSave Text File Print Data on PrinterTo print data To save data as textTo edit a plot label Edit Plot LabelsHOT Plot Status BAR To Magnify a portion of the Data Graph Magnify ToolMagnify MAP Magnified Area Axis Labels To select a channel to graph Channel DataTo link plot lines Channels 4 & 5 After Linking Data Graph ∙ X-Cursor Data Table ∙ Data Plots∙ X-Data Cursors To activate the Help menu Help MenuTo quit List Existing LOG Files on PrinterTo list log files Appendix a KEY Functions