O P E R A T I O N
General Operating Procedures
General Operating Procedures
C H A P T E R 2
The following sections present some guidelines for operation, as well as some common operating procedures.
Warm-Up and Environmental Considerations
Operate the
Temperature Mode Operation
You can operate the
aPlug the
bTurn on the
cPress the ENBL switch in the ADJUST section of the front panel so that the indicator is lit (adjustment enabled). Press the MODE switch until the T mode is selected.
dCheck the setting of the SENSOR SELECT switch for the desired operation. The sensor code will be displayed for two seconds during the
ePress the PARAMETER switch and check the setting of I LIMIT, T LIMIT, and GAIN. Press the CAL switch and check the setting of C1 and C2 to insure that they are compatible with the equipment you are using. Refer to Section 2.4.5 if you need to change them.
fPress the DISPLAY switch until the VIEW SET indicator is lit and check the set point temperature. If it requires changing, turn the knob until the desired value is displayed.
Note: In some cases, a greater than 0.1oC temperature set point resolution may be attained by using R mode with the appropriate resistance value.
gTurn the TEC output on by pressing the OUTPUT ON switch. The unit will automatically control the temperature to the set point.
hWhen the unit is powered off, the state of the unit at
06_07 | 15 |