C H A P T E R 3

MAINTENANC E AND TR OUBLESHOOTING

Calibration Adjustments

AD590 Sensor Calibration

The following procedure is for calibrating the AD590 sensor measurement so that the temperature measurement will be as accurate as possible. This procedure calibrates the current measurement of the AD590. This procedure does not calibrate C1 and C2. For information on calibrating the AD590 sensor, see Appendix C.

Calibration may be aborted by pressing the OUTPUT switch.

aSet the SENSOR SELECT switch (back panel) to the AD590 position. Set C1 to 0.00, C2 to 1.00.

bConnect a precision 20 KΩ metal film resistor and a precision ammeter in series at the sensor input of the LDT-5525 Temperature Controller.

cEnter the sensor calibration mode by pushing the MODE and SENSOR CAL switches at the same time. After this, the TEC display will indicate sensor reference current in µA. Wait for three seconds for the measurement to settle.

dPress and hold in the ENBL switch and turn the ADJUST knob until the display indicates the same current as shown on the precision ammeter.

eRelease the ENBL and wait for the VIEW SET indicator (LED) to be unlit. Replace the 20 KΩ resistor with a 10 KΩ metal film resistor. Wait for three seconds, then repeat Step d using the 10 KΩ resistor.

Ten seconds after the ENBL switch is released, the LDT-5525 Temperature Controller will store the calibration data in non-volatile memory.

fPress the DISPLAY switch three times to rotate the display back to the SENSOR mode. After calibration, I LIMIT will be automatically set to 4.00 Amps. Reset the I LIMIT to the desired value.

22 LDT-5525

Image 36
Contents User’s Guide Page Table of Contents Appendix a Steinhart-Hart Equation Appendix C AD590 and LM335 Sensor Calibration Iv LDT-5525 List of Figures Vi LDT-5525 Safety Information and the Manual General Safety ConsiderationsSafety Symbols Safety Marking SymbolsWarranty Claims for Shipping Damage Comments, Suggestions, and Problems When you contact us, please have the following informationXii LDT-5525 Introduction and Specifications Product OverviewAvailable Options and Accessories Description Model NumberSpecifications Output1User Calibration General Specifications LDT-5525 AC Power Considerations OperationRack Mounting Power-Up Sequence Introduction to the LDT-5525 Front PanelAdjustments Sensor CAL Output Sensor Select Switch Back Panel Controls and ConnectionsTEC Connector Back Panel TEC ConnectorTEC Grounding Considerations General Operating Procedures Temperature Mode OperationWarm-Up and Environmental Considerations Resistance Mode Operation External Safety Switch OperationBooster Operation General Operating Procedures LDT-5525 Maintenance and Troubleshooting Calibration OverviewRecommended Equipment Environmental ConditionsWarm Up Calibration Adjustments Thermistor CalibrationAD590 Sensor Calibration LM335 Sensor Calibration ITE Current Calibration Troubleshooting Symptom Causes and Corrective ActionsTemperature, based on C1 and C2 STEINHART-HART Equation Table A.1 Comparison of Curve Fitting EquationsThermistor R/T Curves Computer Program REM Print Using H$ I, TI, RI LDT-5525 Sensing Current Thermistor Selection Thermistor RangeTable B.2 10K Thermistor Temperature Range Temperature Resolution Selecting and Using ThermistorsSelecting the Sensing Current Thermistor Range AD590 and LM335 Sensor Calibration AD590 SensorLM335 Sensor Figure C.1 AD590 Nonlinearity0607 One Point Calibration Method Two Point Calibration MethodC1n = U + V * C1 and C2n = V * C2 LDT-5525