Lightwave Communications LDT-5525 manual Recommended Equipment, Environmental Conditions, Warm Up

Page 34

C H A P T E R 3

MAINTENANC E AND TR OUBLESHOOTING

Calibration Overview

Recommended Equipment

Recommended test equipment for calibrating the LDT-5525 Temperature Controller is listed in Table 3.1. Equipment other than that shown in the table may be used if the specifications meet or exceed those listed.

Table 3.1

DESCRIPTION

MFG / MODEL

SPECIFICATION

 

 

 

DMM

HP 3457A

DC Amps (@ 1.0A): +0.02%

 

 

Resistance (@ 10Ω): 0.02%

 

 

0.1 µA or 0.1 mW resolution

Resistors

Metal Film

15 KΩ (for ITE calibration)

 

 

4

KΩ and 40 KΩ (for 100 mA calibration)

 

 

4

KΩ and 400 KΩ (for 10 mA calibration)

 

 

4

KΩ and 10 KΩ (for LM335 calibration)

 

 

10 KΩ and 20 KΩ (for AD590 calibration)

 

High Power

1

Ω, 20W, low TCR (for ITE calibration)

 

 

 

 

Environmental Conditions

Calibrate this instrument under laboratory conditions. We recommend calibration at 23oC ± 1.0oC. When necessary, however, the LDT-5525 Temperature Controller may be calibrated at its intended use temperature if this is within the specified operating temperature range of 0 to 40oC.

Warm Up

The LDT-5525 Temperature Controller should be allowed to warm up for at least 1 hour before calibration.

20 LDT-5525

Image 34
Contents User’s Guide Page Table of Contents Appendix a Steinhart-Hart Equation Appendix C AD590 and LM335 Sensor Calibration Iv LDT-5525 List of Figures Vi LDT-5525 Safety Information and the Manual General Safety ConsiderationsSafety Symbols Safety Marking SymbolsWarranty Claims for Shipping Damage Comments, Suggestions, and Problems When you contact us, please have the following informationXii LDT-5525 Introduction and Specifications Product OverviewAvailable Options and Accessories Description Model NumberSpecifications Output1User Calibration General Specifications LDT-5525 Operation AC Power ConsiderationsRack Mounting Power-Up Sequence Introduction to the LDT-5525 Front PanelAdjustments Sensor CAL Output Sensor Select Switch Back Panel Controls and ConnectionsTEC Connector Back Panel TEC ConnectorTEC Grounding Considerations Temperature Mode Operation General Operating ProceduresWarm-Up and Environmental Considerations Resistance Mode Operation External Safety Switch Operation Booster Operation General Operating Procedures LDT-5525 Maintenance and Troubleshooting Calibration OverviewEnvironmental Conditions Recommended EquipmentWarm Up Calibration Adjustments Thermistor CalibrationAD590 Sensor Calibration LM335 Sensor Calibration ITE Current Calibration Troubleshooting Symptom Causes and Corrective ActionsTemperature, based on C1 and C2 STEINHART-HART Equation Table A.1 Comparison of Curve Fitting EquationsThermistor R/T Curves Computer Program REM Print Using H$ I, TI, RI LDT-5525 Sensing Current Thermistor Selection Thermistor RangeTable B.2 10K Thermistor Temperature Range Selecting and Using Thermistors Temperature ResolutionSelecting the Sensing Current Thermistor Range AD590 and LM335 Sensor Calibration AD590 SensorLM335 Sensor Figure C.1 AD590 Nonlinearity0607 One Point Calibration Method Two Point Calibration MethodC1n = U + V * C1 and C2n = V * C2 LDT-5525