C H A P T E R 3
MAINTENANC E AND TR OUBLESHOOTING
Calibration Overview
Recommended Equipment
Recommended test equipment for calibrating the
Table 3.1
DESCRIPTION | MFG / MODEL | SPECIFICATION | |
|
|
| |
DMM | HP 3457A | DC Amps (@ 1.0A): +0.02% | |
|
| Resistance (@ 10Ω): 0.02% | |
|
| 0.1 µA or 0.1 mW resolution | |
Resistors | Metal Film | 15 KΩ (for ITE calibration) | |
|
| 4 | KΩ and 40 KΩ (for 100 mA calibration) |
|
| 4 | KΩ and 400 KΩ (for 10 mA calibration) |
|
| 4 | KΩ and 10 KΩ (for LM335 calibration) |
|
| 10 KΩ and 20 KΩ (for AD590 calibration) | |
| High Power | 1 | Ω, 20W, low TCR (for ITE calibration) |
|
|
|
|
Environmental Conditions
Calibrate this instrument under laboratory conditions. We recommend calibration at 23oC ± 1.0oC. When necessary, however, the
Warm Up
The
20