Lightwave Communications LDT-5525 manual ITE Current Calibration

Page 38

C H A P T E R 3

MAINTENANC E AND TR OUBLESHOOTING

Calibration Adjustments

ITE Current Calibration

The following procedure is for calibrating the ITE constant current source for both polarities of current. During this procedure the ITE current is driven to a series of pre-determined values. When each of these values is reached and is stable, the user enters the actual value of the current, as measured by an external DMM. The LDT-5525 Temperature Controller then automatically calibrates the TEC current source and limits.

Calibration may be aborted by pressing the OUTPUT switch.

aSet the sensor select (back panel) switch to "100 uA." Set C1 to 0.99, C2 to 2.57. Connect an (approximately) 15 kΩ resistor to the sensor pins (7 and 8). Connect a 1Ω, 20 W, resistor across the TEC output terminals (pins 1 and 3) and use a calibrated DMM to measure the voltage across the resistor. Calculate the current in the following steps by using Ohm's Law:

I = E / R

-where E is the accurately measured voltage across the resistor, and R is the accurately measured load resistance. A 4-point probe resistance measurement is recommended.

bPress and breifly hold in both the MODE and DISPLAY switches. This will put the LDT-5525 into ITE calibration mode. Wait for three seconds for the output to settle to about 3 Amps.

cPress and hold in the ENBL switch and turn the ADJUST knob until the display shows the correct value (absolute value of the ITE measurement), as calculated from Step a.

dRelease the ENBL switch. Wait three seconds to allow the ITE current to settle at the new set point.

eRepeat Steps c and d four more times, once for each of the (automatically set) set points -3 Amps, +3 Amps, +1 Amp, and -1 Amp. After the value for the -1 Amp (last) set point is entered, the LDT-5525 Temperature Controller will automatically calibrate its ITE current limits.

After about 10 seconds, the LDT-5525 Temperature Controller will store the new calibration data in non-volatile memory.

fAfter calibration, the I LIMIT will be automatically set to 0.00 Amps. Reset the I LIMIT to the desired value.

24 LDT-5525

Image 38
Contents User’s Guide Page Table of Contents Appendix a Steinhart-Hart Equation Appendix C AD590 and LM335 Sensor Calibration Iv LDT-5525 List of Figures Vi LDT-5525 Safety Information and the Manual General Safety ConsiderationsSafety Symbols Safety Marking SymbolsWarranty Claims for Shipping Damage Comments, Suggestions, and Problems When you contact us, please have the following informationXii LDT-5525 Introduction and Specifications Product OverviewAvailable Options and Accessories Description Model NumberSpecifications Output1User Calibration General Specifications LDT-5525 Rack Mounting AC Power ConsiderationsOperation Power-Up Sequence Introduction to the LDT-5525 Front PanelAdjustments Sensor CAL Output Sensor Select Switch Back Panel Controls and ConnectionsTEC Connector Back Panel TEC ConnectorTEC Grounding Considerations Warm-Up and Environmental Considerations General Operating ProceduresTemperature Mode Operation Resistance Mode Operation External Safety Switch OperationBooster Operation General Operating Procedures LDT-5525 Maintenance and Troubleshooting Calibration OverviewWarm Up Recommended EquipmentEnvironmental Conditions Calibration Adjustments Thermistor CalibrationAD590 Sensor Calibration LM335 Sensor Calibration ITE Current Calibration Troubleshooting Symptom Causes and Corrective ActionsTemperature, based on C1 and C2 STEINHART-HART Equation Table A.1 Comparison of Curve Fitting EquationsThermistor R/T Curves Computer Program REM Print Using H$ I, TI, RI LDT-5525 Sensing Current Thermistor Selection Thermistor RangeTable B.2 10K Thermistor Temperature Range Selecting the Sensing Current Temperature ResolutionSelecting and Using Thermistors Thermistor Range AD590 and LM335 Sensor Calibration AD590 SensorLM335 Sensor Figure C.1 AD590 Nonlinearity0607 One Point Calibration Method Two Point Calibration MethodC1n = U + V * C1 and C2n = V * C2 LDT-5525