C H A P T E R 3
MAINTENANC E AND TR OUBLESHOOTING
Calibration Adjustments
ITE Current Calibration
The following procedure is for calibrating the ITE constant current source for both polarities of current. During this procedure the ITE current is driven to a series of
Calibration may be aborted by pressing the OUTPUT switch.
aSet the sensor select (back panel) switch to "100 uA." Set C1 to 0.99, C2 to 2.57. Connect an (approximately) 15 kΩ resistor to the sensor pins (7 and 8). Connect a 1Ω, 20 W, resistor across the TEC output terminals (pins 1 and 3) and use a calibrated DMM to measure the voltage across the resistor. Calculate the current in the following steps by using Ohm's Law:
I = E / R
bPress and breifly hold in both the MODE and DISPLAY switches. This will put the
cPress and hold in the ENBL switch and turn the ADJUST knob until the display shows the correct value (absolute value of the ITE measurement), as calculated from Step a.
dRelease the ENBL switch. Wait three seconds to allow the ITE current to settle at the new set point.
eRepeat Steps c and d four more times, once for each of the (automatically set) set points
After about 10 seconds, the
fAfter calibration, the I LIMIT will be automatically set to 0.00 Amps. Reset the I LIMIT to the desired value.
24