MAINTENANCE AND TR OUBLESHOOTING

Calibration Adjustments

LM335 Sensor Calibration

C H A P T E R 3

The following procedure is for calibrating the LM335 sensor measurement so that the temperature measurement will be as accurate as possible. This procedure calibrates the voltage measurement of the LM335. This procedure does not calibrate C1 and C2. For information on calibrating the LM335 sensor, see Appendix C.

Calibration may be aborted by pressing the OUTPUT switch.

aSet the SENSOR SELECT switch (back panel) to the LM335 position. Set C1 to 0.00, C2 to 1.00.

bConnect a precision 4 KΩ metal film resistor and a precision voltmeter in parallel at the sensor input of the LDT-5525 Temperature Controller (pins 7 and 8).

cEnter the sensor calibration mode by pushing the MODE and SENSOR CAL switches at the same time. After this, the display will indicate sensor reference voltage in mV. Wait for three seconds for the measurement to settle.

dPress and hold in the ENBL switch and turn the ADJUST knob until the display indicates the same voltage as shown on the precision voltmeter multiplied by 10. For example, if the voltage across the resistor is 1.9871 Volts, turn the ADJUST knob until the display reads 19.87.

eRelease the ENBL switch and wait for the VIEW SET indicator (LED) to be unlit. Replace the 4 KΩ resistor with a 10 KΩmetal film resistor. After three seconds, repeat Step d with the 10 K* resistor. Ten seconds after the ENBL switch is released, the LDT-5525 Temperature Controller will store the calibration data in non-volatile memory.

fPress the DISPLAY switch three times to rotate the display back to the SENSOR mode. After calibration, I LIMIT will be automatically set to 4.00 Amps. Reset the I LIMIT to the desired value.

06_07

LDT-5525

23

Image 37
Contents User’s Guide Page Table of Contents Appendix a Steinhart-Hart Equation Appendix C AD590 and LM335 Sensor Calibration Iv LDT-5525 List of Figures Vi LDT-5525 General Safety Considerations Safety Information and the ManualSafety Marking Symbols Safety SymbolsWarranty Claims for Shipping Damage When you contact us, please have the following information Comments, Suggestions, and ProblemsXii LDT-5525 Product Overview Introduction and SpecificationsDescription Model Number Available Options and AccessoriesOutput1 SpecificationsUser Calibration General Specifications LDT-5525 Operation AC Power ConsiderationsRack Mounting Introduction to the LDT-5525 Front Panel Power-Up SequenceAdjustments Sensor CAL Output Back Panel Controls and Connections Sensor Select SwitchBack Panel TEC Connector TEC ConnectorTEC Grounding Considerations Temperature Mode Operation General Operating ProceduresWarm-Up and Environmental Considerations External Safety Switch Operation Resistance Mode OperationBooster Operation General Operating Procedures LDT-5525 Calibration Overview Maintenance and TroubleshootingEnvironmental Conditions Recommended EquipmentWarm Up Thermistor Calibration Calibration AdjustmentsAD590 Sensor Calibration LM335 Sensor Calibration ITE Current Calibration Symptom Causes and Corrective Actions TroubleshootingTemperature, based on C1 and C2 Table A.1 Comparison of Curve Fitting Equations STEINHART-HART EquationThermistor R/T Curves Computer Program REM Print Using H$ I, TI, RI LDT-5525 Thermistor Range Sensing Current Thermistor SelectionTable B.2 10K Thermistor Temperature Range Selecting and Using Thermistors Temperature ResolutionSelecting the Sensing Current Thermistor Range AD590 Sensor AD590 and LM335 Sensor CalibrationFigure C.1 AD590 Nonlinearity LM335 Sensor0607 Two Point Calibration Method One Point Calibration MethodC1n = U + V * C1 and C2n = V * C2 LDT-5525