Troubleshooting

 

Unable to adjust output or

Check the ADJUST ENBL switch; the indicator must be lit for any

 

 

parameter

adjustments to be made

 

 

 

Check the MODE or DISPLAY switch; if they do not respond, the

 

 

 

unit may be in measurement calibration mode (see Chapter 3);

 

 

 

press the OUTPUT switch to abort this mode.

 

 

Unable to switch DISPLAY,

The unit may be in measurement calibration mode (see Chapter

 

 

MODE, SENSOR CAL or

3); press the OUTPUT switch to abort this mode

 

 

PARAMETER modes

 

 

 

 

 

 

 

Output goes off intermittently

Check that the AC power cord connection is secure; power line

 

 

 

drop outs may reset the unit and when the power is restored, the

 

 

 

output will be off

 

 

 

Check the TE module connections; a high impendance on the TE

 

 

 

load may cause the output to exceed the compliance voltage

 

 

 

momentarily, thus shutting the output off

 

 

R Mode set point is not saved

The R mode set point value is not independent from the T mode

 

 

 

set point value; if the control mode is changed from R mode to T

 

 

 

mode, the R set value will change to a value which corresponds to

 

 

 

the temperature, based on C1 and C2

 

 

 

 

 

26LDT-5525

Image 40
Contents User’s Guide Page Table of Contents Appendix a Steinhart-Hart Equation Appendix C AD590 and LM335 Sensor Calibration Iv LDT-5525 List of Figures Vi LDT-5525 Safety Information and the Manual General Safety ConsiderationsSafety Symbols Safety Marking SymbolsWarranty Claims for Shipping Damage Comments, Suggestions, and Problems When you contact us, please have the following informationXii LDT-5525 Introduction and Specifications Product OverviewAvailable Options and Accessories Description Model NumberSpecifications Output1User Calibration General Specifications LDT-5525 Operation AC Power ConsiderationsRack Mounting Power-Up Sequence Introduction to the LDT-5525 Front PanelAdjustments Sensor CAL Output Sensor Select Switch Back Panel Controls and ConnectionsTEC Connector Back Panel TEC ConnectorTEC Grounding Considerations Temperature Mode Operation General Operating ProceduresWarm-Up and Environmental Considerations Resistance Mode Operation External Safety Switch OperationBooster Operation General Operating Procedures LDT-5525 Maintenance and Troubleshooting Calibration OverviewEnvironmental Conditions Recommended EquipmentWarm Up Calibration Adjustments Thermistor CalibrationAD590 Sensor Calibration LM335 Sensor Calibration ITE Current Calibration Troubleshooting Symptom Causes and Corrective ActionsTemperature, based on C1 and C2 STEINHART-HART Equation Table A.1 Comparison of Curve Fitting EquationsThermistor R/T Curves Computer Program REM Print Using H$ I, TI, RI LDT-5525 Sensing Current Thermistor Selection Thermistor RangeTable B.2 10K Thermistor Temperature Range Selecting and Using Thermistors Temperature ResolutionSelecting the Sensing Current Thermistor Range AD590 and LM335 Sensor Calibration AD590 SensorLM335 Sensor Figure C.1 AD590 Nonlinearity0607 One Point Calibration Method Two Point Calibration MethodC1n = U + V * C1 and C2n = V * C2 LDT-5525