ELECTROSTATIC DISCHARGE (ESD)
The SSDs will meet the ESD limits specified in the
| Table 43. | ESD Requirements | |
|
| ||
| Climatic Conditions | ||
|
|
|
|
Ambient Temperature | 15oC to 35oC |
|
|
Relative Humidity | 30% to 60% |
|
|
|
| ||
Atmospheric Pressure | 86kPa (860 millibar) to 106 kPa (1,060 millibar) | ||
|
|
|
|
Voltage Level | Discharge Type |
| Pass/Fail Performance Criteria |
|
|
|
|
Contact |
| A | |
|
|
|
|
Contact |
| B | |
|
|
|
|
Contact |
| B | |
|
|
|
|
Contact |
| C | |
|
|
|
|
Air |
| A | |
|
|
|
|
Air |
| A | |
|
|
|
|
Air |
| B | |
|
|
|
|
Air |
| B | |
|
|
|
|
Air |
| C | |
|
|
|
|
ACCEPTANCE CRITERIA DEFINITIONS
The following table lists the acceptance criteria definitions for the ESD limits.
Table 44. Acceptance Criteria Definitions
AThe apparatus will continue to operate as intended, i.e., normal unit operation with no degradation of performance.
The apparatus will continue to operate as intended after completion of the test. However,
Bduring the test, some degradation of performance is allowed, provided there is no data lost or operator intervention to restore apparatus function.
CTemporary loss of function is allowed. Operator intervention is acceptable to restore apparatus function.
Note: Hardware failures are not acceptable for any level of the above performance criteria.
76 | ZeusIOPS |