Press Done (F2) to exit the editing screen.
Printing or Downloading Test Results
To send test results directly from the Scanner to a printer or PC:
1.Configure the Printer Port settings in the PRINTER PORT SETUP screen. To display this screen on the PentaScanner or PentaScanner Cable Admin, press Extended Functions key, then select Printer Port. To display this screen on the PentaScanner+, press the Extended Functions key, then select User Preferences, then select Printer Port.
1.Press the or ∇ arrow key to highlight a setting to modify — Baud rate or Handshaking. The default settings for Baud rate and Handshaking are 9600 and XON/XOFF. The settings for Data Bits, Stop Bits and Parity are fixed at 8, 1 and None and cannot be changed. After highlighting the Baud rate or Handshaking setting, press Change (F2) to toggle through the available settings.
2. Press Printer (F3) to display the | |||||
graphics printer screen. Press Save (F1) | |||||
|
|
|
| ||
|
|
|
| to store the new settings and exit or press | |
|
|
|
| the ESCAPE key to exit without saving. | |
|
|
|
| 3. Choose the appropriate serial | |
|
|
|
| communications cable. A | |
|
|
|
| serial cable is included in the kit, or you | |
|
|
|
| can use a | |
Male | Female | Male | Female | The cable must be wired as in Appendix | |
C: | |||||
|
|
|
|
1.Attach the
Connection for Printing Test Results
Printer (rear view)
ScannerDB-9
Serial Port
Serial Cable
Connection for Downloading Test Results
COM 1 or COM 2
Scanner
Serial Cable
1.Use the PRINT OPTIONS screen to set the criteria to determine which Autotests to print. To display the PRINT OPTIONS screen, press the Extended Functions key, then select
Memory, then press Print (F3).
2.To display the PRINT FORMAT screen, press Print (F3) from the AUTOTEST RESULTS screen, or press Print (F3) from the PRINT OPTIONS screen. Press Change (F2) from the
Page 33