be calibrated each calendar day that it is used. Expiration of calibration data is automatically detected by the Scanner.
The Scanner and Injector are connected through their respective test connector ports using the patch cable provided with each instrument. The amplitude of each frequency generated by the Injector is measured by the Scanner. The resulting calibration values are stored in the Scanner according to the Injector’s serial number. The Scanner maintains five sets of Injector calibration data simultaneously.
1.Connect the Scanner and the Injector using the patch cable supplied with the Scanner (see following diagrams). If you have a Scanner with a
2.Power on the Scanner.
3.Press the Extended Functions key.
4.Select Calibrate Injector. The Scanner will display: Searching for Injector . . .
5.On finding the Injector, the Scanner displays: Calibrating . . .
6.When the calibration is complete, the Scanner displays: The Injector has been calibrated
7.Press OK (F1) to return to the Extended Functions menu.
Mod 8 Adapter Note
Use of the mating connectors not in conformance with FCC Part 68,
Calibration Connections
Modular 8 Adapter
Modular 8 Plug to
Page 9