NumaLink-3.0™ Users Guide and Service Manual

Figure 15 – NumaLink-3.0 Pegasys Write Configuration

Numa System

ADAC Pegasys

NumaLink

AutoFTP

 

1

 

2

 

 

 

2

3

 

local

local

 

disk

disk

 

 

NL_UserMan_003

1.Completing a translation, NumaLink-3.0 writes Pegasys files to a configured output directory.

2.AutoFTP transfers the Pegasys files to a local directory on the Pegasys system.

3.AutoFTP shells a procedure on the Pegasys to add the transferred file to the Pegasys database.

6.2 GE StarCam

6.2.1 NumaLink-3.0 Input Configuration (StarCam Read)

To transfer patient studies from your GE StarCam to the NumaLink-3.0 system:

1.At the > prompt on the StarCam console, type the command NSND and press the Enter key. This will bring up the Network Send screen used to select a station and patient.

2.Press the Help key to bring up a list of available StarLink stations. Select the entry for the NumaLink system, (This entry name will vary, it may be “numa” or “adac” or “odyssey” – contact Numa if you need assistance) then press the SEL key to select the entry.

3.Press the Enter key to navigate to the Patient ID field. Press the Help key to bring up a patient list. Using the arrow keys, select the patient you wish to transfer and press the SEL key to select the patient. This will bring you back to the main Network Send screen.

4.To send the patient, press the Send Form key.

6.3 Marconi Odyssey

6.3.1 NumaLink-3.0 Input Configuration (Odyssey Read)

To transfer patient studies from your Marconi Odyssey computer to the Numa computer:

Page 25 of 33

Image 26
Contents NumaLink-3.0 Version AprilFor more information contact Numa Table of Contents Introduction Cabling Installation and Getting StartedHardware Shutdown and Reboot StartupLicense Installation DirectoriesSoftware License Error Message Removing NumaLink-3.0Product Overview NumaLinkTypical Translation DicomPage Configuration GeneralNumaLInk-3.0 Dicom Application Entity Information Translations Translations Tab Translations Settings Customization DicomDicom dialog Dicom Settings Status NumaLink-3.0 Dicom Conformance StatementPage Technical Support Logging and Error ReportingRemote Diagnostics Troubleshooting PcAnywhere IconFrequently Asked Questions Page Page NumaLink-3.0 Output Configuration Pegasys Write OEM Setup InformationNumaLink-3.0 Input Configuration Pegasys Read Adac PegasysGE StarCam NumaLink-3.0 Input Configuration StarCam ReadNumaLink-3.0 Input Configuration Odyssey Read Marconi OdysseyNumaLink-3.0 Input Configuration Icon Read Siemens IconCommand Line Interface and Configuration File Command LineConfiguration File Translate Directive Instead Outremoteae Directive Control* Directives