Agilent Technologies Understanding Service Requests for Agilent 6623A Power Supply

Page 73

To find out the nature of the service request, you must do a serial poll. This will isolate the output that generated the request by checking which of the FAU bits are set in the case of a fault, or checking to see if the error bit is set in the case of an error. If the SRQ on faults was set, then send the fault query.

FAULT? 2 (using output 2 as an example)

and address the supply to talk if you want to find out which of the conditions you unmasked in Figure 5-3 are true. For example if the supply was in overvoltage and that condition was unmasked then the response from the fault query will be ’’8" (see Table 5-5).

NOTE

When you query the fault, the fault register is cleared. Performing a serial poll will reset the PQS bit but

 

will not clear the fault register.

 

 

If the SRQ on error was set, then you can send the error query ERR? and address the supply to talk. The response will identify the error by its code (see Table 5-8).

Service Request Enable/Disable. You can query the status of the service request enable/disable function by sending the query:

SRQ?

and addressing the power supply to talk. The response from the supply is one of the following:

0, 1, 2, or 3

0--indicates that the service request capability (except for power-on; see The Power On-Service Request information below) is disabled.

1--indicates that it is enabled for output fault conditions.

2--indicates that it is enabled for error conditions.

3--indicates that it is enabled for both fault and error conditions

The ability to generate service requests can be enabled or disabled using the SRQ command as described below.

To disable the service request capability, except for power-on, send:

SRQ 0

To enable the service request capability for all output faults,

SRQ 1

To enable the service request capability for errors, send:

SRQ 2

To enable the service request capability for both faults and errors, send:

SRQ 3

The Power-On Service Request. You can also cause the power supply to request service every time it is switched on or every time there is a temporary loss in power. To do this send the following command:

PON 1

76 Remote Operation

Image 73
Contents Agilent Part No Operating ManualCertification Safety Summary Environmental Conditions Safety SummaryEMC Declaration of ConformityWhat this Manual Contains Table Of Contents Local Operation Remote OperationError Messages Command SummaryCalibration Programming With a Series 200/300 ComputerGeneral Information Safety ConsiderationsIntroduction Instrument and Manual IdentificationDescription AccessoriesModel Output Combinations AvailableGP-IB Board Basic OperationOutput Boards Qualifying Conditions SpecificationsDefinitions Output Response Characteristics Source Effect SpecificationsOutputs Low High Voltage Temperature Coefficient Supplemental CharacteristicsOVP Readback ResolutionLow Voltage General Information General Information General Information General Information Initial Inspection InstallationLocation and Cooling Line Fuse Input Power RequirementsLine Fuses GP-IBLine Voltage Conversion Power CordGP-IB Interface Connector Getting Started Front Panel Controls and IndicatorsTurning On Your Supply 15V 35A Output Controls and Indicators Number Controls/lndicators Normal Self Test Indications Test Pattern of all Display Segments at Power-onSample Self-Test Failure Display Checking Out Your Supply Using Local ControlOvervoltage Test Voltage TestCurrent Test Iset Enter Introduction To Remote OperationOCP RSTOutput Sending a Remote CommandReading the GP-IB Address AddrGetting Data From The Supply Often Used CommandsDisp a Disp a Returning the Supply to Local Mode Output Ranges Output Connections and Operating InformationOperating Quadrants Protection FeaturesRange Selection Typical Output Range Characteristics Connecting the Load Page AWG Wire Size Wire Bundled 10 a 20 aMultiple Loads Remote Voltage SensingRemote Sense Connections Remote Voltage SensingOutput Noise Considerations Output Type FormulaProgramming Response Time with an Output Capacitor Open Sense LeadsExternal Trigger Circuit Overvoltage Trigger ConnectionsEquivalent Internal OV Trigger Circuit Battery Charging Power Supply Protection ConsiderationsParallel Operation CV Operation Maximum Allowable Voltage SettingRemote Sensing CC Operation13. Series Connections with Local Sensing CV Operation Series Operation14. Series Connections with Remote Sensing Specifications for Series OperationPage GP-IB Operation Remote OperationInterface Function GP-IB Address Selection Programming Syntax Power-On Service Request PONNumeric Data Sheet 1 of 2. Syntax Forms for Power Supply Commands Sheet 2 of 2. Syntax Forms for Power Supply Commands Data Range Power Supply Commands Header Output ChannelInitial Conditions Power Supply CommandsVSET? Voltage ProgrammingVOUT? Current ProgrammingIOUT? Avg Current-Avg RangeAvg ResolutionOutput On/Off Range SwitchingOVSET? Overvoltage OV ProtectionOvercurrent Protection OCP Multiple Output Storage & Recall Clear CommandStatus Reporting UNR +CC Functional Relationship of Status RegistersUNMASK? ASTS?Unmask 2,XXX Service Request Generation FAULT?Bit Assignment of the Serial Poll Register PON RQS ERR RDY FAUSRQ? RQS Bit Reprogramming DelayOther Queries Display On/OffCMODE? TEST?Front Panel Response GP-IB Code Error Messages ExplanationResponse Code Front PanelTEST? Responses Code ExplanationLocal Operation Local ModeLocal Control Of Output Functions GeneralSetting Current Setting VoltageResetting Overvoltage Protection Setting Overvoltage ProtectionResetting Overcurrent Protection Displaying the Contents of the Fault RegisterSetting the Supply’s GP-IB Address Setting the Reprogramming DelayLocal Control Of System Functions ConditionAddr Enter Displaying Error MessagesSTO Enter RCL EnterTest Equipment and Setup Required Calibration ProceduresFigure A-1. Calibration Setup Table A-1. Calibrat ion Commands Header Channel Data Syntax General Calibration ProcedureSee Figure Page 10 ! Calibration Example Calibration ProgramClear Voltmeter Output Buffer PauseDisp END of Calibration Program Input ANY More Outputs to CALIBRATE? Y or N,X$Fnend Page Path Names Programming With a Series 200/300 ComputerVoltage and Current Programming Voltage and Current Readback Voltage and Current Programming With VariablesPrint OUTPUT1 is in CV Mode END if Programming Power Supply RegistersService Request and Serial Poll Present StatusOFF Intr Enable IntrPrint ’’OVERVOLTAGE on Output #1 Print Overvoltage on Output #2Error Detection Stored Operating States Programming Outputs Connected In ParallelInput Enter Operating VOLTAGE,V1 Input Enter Voltage LIMIT’’,VInput Enter the Desired Current Limit POINT,I Programming Outputs Connected In SeriesTable C-1. Command Summary Command SummaryCommand Description Table C-l. Command Summary ROM? PON?SRQ? Power-On Self Test Messages Error Responses Error Codes and MessagesTable D-l. Power-On Self Test Error Message Test ResponsesError Code Message Explanation ERR? query ERR key Table D-2. Error ResponsesResponse Code Explanation TEST? query Table D-3. TEST? ResponsesMake Changes Manual BackdatingII. CE’92 Product Specific Annotations Generally Applicable Annotations6621A 6623AUnited States Latin America Agilent Sales and Support OfficeManual Updates

6627A, 6621A, 6624A, 6623A, 6622A specifications

Agilent Technologies is renowned for its high-quality electronic test and measurement equipment, and the Agilent 6600 series is no exception. This series includes models like the Agilent 6621A, 6622A, 6623A, 6624A, and 6627A, each designed to meet the needs of various application requirements, making them an essential part of modern laboratories.

The Agilent 6621A is a single-output DC power supply that provides a stable output voltage and current, making it ideal for testing and powering electronic devices. It features a low noise specification, which is crucial for sensitive applications. With a maximum output voltage of 30V and a current of 3A, it offers flexibility for a range of projects, from powering prototypes to performing benchmark tests.

The Agilent 6622A, a dual-output model, enhances versatility by allowing users to power two devices concurrently. It delivers output voltages of up to 20V and a total output current of 5A, which is perfect for powering circuit boards with multiple components. The built-in voltage and current limiting functions protect the equipment under test, preventing any potential damage.

On the other hand, the Agilent 6623A provides additional capabilities with its three outputs, making it particularly suitable for complex testing procedures. With a maximum voltage of 20V and output current reaching 6A across all channels, it ensures that multiple loads can be powered simultaneously without compromising performance.

The Agilent 6624A further pushes these capabilities with its higher output power. This model boasts two outputs with a combined maximum output of up to 6A, supporting devices that require more demanding power levels. Its advanced control features allow for precise voltage and current adjustments, enhancing reliability during experiments.

Lastly, the Agilent 6627A stands out as a highly scalable power supply, capable of delivering up to 40V and 7.5A across its multiple outputs. This model is particularly beneficial for applications requiring higher voltages, enabling engineers and technicians to work with a broader array of components and systems.

All models in the Agilent 6600 series incorporate built-in protection features to guarantee safety during testing. They are equipped with memory functions, allowing users to save and recall settings quickly. Additionally, the intuitive interface and various connectivity options make these power supplies user-friendly, ensuring efficient workflow in any laboratory setting. In summary, the Agilent 6600 series offers a compelling combination of versatility, precision, and advanced features, catering to diverse electronic testing applications.