Agilent Technologies 6624A, 6621A, 6627A, 6622A Table C-1. Command Summary, Command Description

Page 99

C

Command Summary

Introduction

Table C-1 provides an alphabetical listing and a brief description of each command that can be sent to the Agilent 6621A- 24A, and 6627A power supplies. All of the commands can be executed remotely over the GP-IB. Many of the commands can also be executed locally from the supply’s front panel as indicated in the table.

Command headers are accepted in upper or in lower case letters although only upper case letters are used in this summary. The brackets < > indicate a number to be sent. Note that <ch> must specify an output channel number from 1 through 4 as applicable. Commas or spaces are required between numbers. Spaces are optional between the command header and the first number. No commas are allowed between header and first number. Use semi-colons between multiple commands sent in one statement.

Chapter 5 gives a complete description (including syntax) of most of the commands listed. Calibration commands are described in Appendix A. Two of the commands listed (ROM? and VMUX?) are described in the Service Manual.

 

Table C-1. Command Summary

Command

Description

ASTS? < ch >

Queries the accumulated status (ASTS) of the specified output channel. The

 

response (integer 0-255) represents the sum of the binary weights of the ASTS

 

register bits (see page 74). The ASTS register is automatically set to the present

 

status after being queried.

CLR

Returns the entire power supply (all outputs) to the power on state except that the

 

supply is not unaddressed and its store/recall registers are not changed (see Clear

 

Command page 73).

CMODE <on/off>

Turns the calibration mode on or off. On/off is a 1 to turn the calibration mode on;

 

a 0 turns it off (see Appendix A).

CMODE?

Queries if the calibration mode is on or off. Response is either a 1 (on) or 0 (off),

 

(see page 79).

*DLY < ch >, < delay >

Sets the reprogramming delay for the specified output channel. This delay is used

 

to mask the CV, + CC, - CC, and UNR status bits from the fault register and the

 

OCP function for the specified delay period (0 to 32 seconds). The delay time is

 

initiated after a VSET, ISET, OUT, RCL, OVRST, or OCRST command is sent

 

(see Reprogramming Delay, page 78).

DLY? < ch >

Queries the present reprogramming delay of the specified output channel (see page

 

78). The response can be a real number from 0.000 to 32 seconds (e.g., .450

 

seconds).

DSP < on/off >

Turns the power supply’s front panel display either on or off (see page 78). On/off

 

equals 1 to turn the display on or a 0 to turn it off.

 

 

* Indicates that the command can be executed from the front panel.

Command Summary 105

Image 99
Contents Agilent Part No Operating ManualCertification Safety Summary Environmental Conditions Safety SummaryEMC Declaration of ConformityWhat this Manual Contains Table Of Contents Local Operation Remote OperationProgramming With a Series 200/300 Computer Command SummaryError Messages CalibrationInstrument and Manual Identification Safety ConsiderationsGeneral Information IntroductionOutput Combinations Available AccessoriesDescription ModelGP-IB Board Basic OperationOutput Boards Specifications Qualifying ConditionsDefinitions Output Response Characteristics Source Effect SpecificationsOutputs Low High Voltage Temperature Coefficient Supplemental CharacteristicsOVP Readback ResolutionLow Voltage General Information General Information General Information General Information Installation Initial InspectionLocation and Cooling Line Fuse Input Power RequirementsLine Fuses GP-IBLine Voltage Conversion Power CordGP-IB Interface Connector Front Panel Controls and Indicators Getting StartedTurning On Your Supply 15V 35A Output Controls and Indicators Number Controls/lndicators Normal Self Test Indications Test Pattern of all Display Segments at Power-onSample Self-Test Failure Display Checking Out Your Supply Using Local ControlVoltage Test Overvoltage TestCurrent Test RST Introduction To Remote OperationIset Enter OCPAddr Sending a Remote CommandOutput Reading the GP-IB AddressOften Used Commands Getting Data From The SupplyDisp a Disp a Returning the Supply to Local Mode Output Ranges Output Connections and Operating InformationProtection Features Operating QuadrantsRange Selection Typical Output Range Characteristics Connecting the Load Page AWG Wire Size Wire Bundled 10 a 20 aMultiple Loads Remote Voltage SensingRemote Sense Connections Remote Voltage SensingOpen Sense Leads Output Type FormulaOutput Noise Considerations Programming Response Time with an Output CapacitorExternal Trigger Circuit Overvoltage Trigger ConnectionsEquivalent Internal OV Trigger Circuit Power Supply Protection Considerations Battery ChargingParallel Operation CV Operation Maximum Allowable Voltage SettingRemote Sensing CC Operation13. Series Connections with Local Sensing CV Operation Series Operation14. Series Connections with Remote Sensing Specifications for Series OperationPage Remote Operation GP-IB OperationInterface Function GP-IB Address Selection Power-On Service Request PON Programming SyntaxNumeric Data Sheet 1 of 2. Syntax Forms for Power Supply Commands Sheet 2 of 2. Syntax Forms for Power Supply Commands Data Range Power Supply Commands Header Output ChannelInitial Conditions Power Supply CommandsCurrent Programming Voltage ProgrammingVSET? VOUT?IOUT? Avg Current-Avg RangeAvg ResolutionOutput On/Off Range SwitchingOvervoltage OV Protection OVSET?Overcurrent Protection OCP Clear Command Multiple Output Storage & RecallStatus Reporting UNR +CC Functional Relationship of Status RegistersASTS? UNMASK?Unmask 2,XXX PON RQS ERR RDY FAU FAULT?Service Request Generation Bit Assignment of the Serial Poll RegisterSRQ? RQS Bit Reprogramming DelayOther Queries Display On/OffCMODE? TEST?Front Panel Response GP-IB Code Error Messages ExplanationCode Explanation Front PanelResponse Code TEST? ResponsesGeneral Local ModeLocal Operation Local Control Of Output FunctionsSetting Current Setting VoltageDisplaying the Contents of the Fault Register Setting Overvoltage ProtectionResetting Overvoltage Protection Resetting Overcurrent ProtectionCondition Setting the Reprogramming DelaySetting the Supply’s GP-IB Address Local Control Of System FunctionsRCL Enter Displaying Error MessagesAddr Enter STO EnterTest Equipment and Setup Required Calibration ProceduresFigure A-1. Calibration Setup General Calibration Procedure Table A-1. Calibrat ion Commands Header Channel Data SyntaxSee Figure Page Pause Calibration Program10 ! Calibration Example Clear Voltmeter Output BufferInput ANY More Outputs to CALIBRATE? Y or N,X$ Disp END of Calibration ProgramFnend Page Programming With a Series 200/300 Computer Path NamesVoltage and Current Programming Voltage and Current Readback Voltage and Current Programming With VariablesPresent Status Programming Power Supply RegistersPrint OUTPUT1 is in CV Mode END if Service Request and Serial PollPrint Overvoltage on Output #2 Enable IntrOFF Intr Print ’’OVERVOLTAGE on Output #1Error Detection Stored Operating States Programming Outputs Connected In ParallelInput Enter Operating VOLTAGE,V1 Input Enter Voltage LIMIT’’,VInput Enter the Desired Current Limit POINT,I Programming Outputs Connected In SeriesCommand Summary Table C-1. Command SummaryCommand Description Table C-l. Command Summary ROM? PON?SRQ? Test Responses Error Codes and MessagesPower-On Self Test Messages Error Responses Table D-l. Power-On Self Test Error MessageError Code Message Explanation ERR? query ERR key Table D-2. Error ResponsesResponse Code Explanation TEST? query Table D-3. TEST? ResponsesMake Changes Manual Backdating6623A Generally Applicable AnnotationsII. CE’92 Product Specific Annotations 6621AUnited States Latin America Agilent Sales and Support OfficeManual Updates

6627A, 6621A, 6624A, 6623A, 6622A specifications

Agilent Technologies is renowned for its high-quality electronic test and measurement equipment, and the Agilent 6600 series is no exception. This series includes models like the Agilent 6621A, 6622A, 6623A, 6624A, and 6627A, each designed to meet the needs of various application requirements, making them an essential part of modern laboratories.

The Agilent 6621A is a single-output DC power supply that provides a stable output voltage and current, making it ideal for testing and powering electronic devices. It features a low noise specification, which is crucial for sensitive applications. With a maximum output voltage of 30V and a current of 3A, it offers flexibility for a range of projects, from powering prototypes to performing benchmark tests.

The Agilent 6622A, a dual-output model, enhances versatility by allowing users to power two devices concurrently. It delivers output voltages of up to 20V and a total output current of 5A, which is perfect for powering circuit boards with multiple components. The built-in voltage and current limiting functions protect the equipment under test, preventing any potential damage.

On the other hand, the Agilent 6623A provides additional capabilities with its three outputs, making it particularly suitable for complex testing procedures. With a maximum voltage of 20V and output current reaching 6A across all channels, it ensures that multiple loads can be powered simultaneously without compromising performance.

The Agilent 6624A further pushes these capabilities with its higher output power. This model boasts two outputs with a combined maximum output of up to 6A, supporting devices that require more demanding power levels. Its advanced control features allow for precise voltage and current adjustments, enhancing reliability during experiments.

Lastly, the Agilent 6627A stands out as a highly scalable power supply, capable of delivering up to 40V and 7.5A across its multiple outputs. This model is particularly beneficial for applications requiring higher voltages, enabling engineers and technicians to work with a broader array of components and systems.

All models in the Agilent 6600 series incorporate built-in protection features to guarantee safety during testing. They are equipped with memory functions, allowing users to save and recall settings quickly. Additionally, the intuitive interface and various connectivity options make these power supplies user-friendly, ensuring efficient workflow in any laboratory setting. In summary, the Agilent 6600 series offers a compelling combination of versatility, precision, and advanced features, catering to diverse electronic testing applications.