Agilent Technologies 82350B manual Installing and Configuring

Page 32

1 Installing and Configuring the 82350

After the 82350 is installed in your PC, the next step is to connect your GPIB instruments to the installed 82350. Suggested steps follow. When you have made the connections, go to “Step 4 - Configuring the 82350”.

1Review Connection Guidelines. The recommended method for connecting a GPIB system is a linear arrangement with the system controller (PC) at one end of the system. However, a GPIB system can also be connected together in a star, linear, or a combination configuration as long as the total number of devices on the system is £ 14 and these guidelines are followed:

To minimize stress on connector mountings, no more than three cable connectors blocks should be stacked on top of one another.

The GPIB connector screws should be finger- tightened only.

Minimize cable length as much as possible. All system devices must have tri- state drivers and must be powered on. Systems with devices not using tri- state drivers are limited to transfer rates < 250 Kbytes/sec.

For operation with data transfer rates < 500

Kbytes/sec, the total length of all GPIB cables is ≤2 meters times the number of

devices connected together, up to a maximum of 20 meters.

For operation with data transfer rates > 500 Kbytes/sec, the total length of all GPIB cables is > 1 meter times the number of devices connected together, up to a maximum of 15 meters.

The length between adjacent devices is not critical as long as the overall restriction is met. GPIB bus extenders are available that allow operation over much greater distances.

20

82350B Installation and Configuration Guide

Image 32
Contents PCI Gpib Manual Part Number Safety Notices Safety Symbols Page General Safety Information Environment Conditions Regulatory Markings Page Declaration of Conformity Contents Specifications and Information Installing and Configuring Getting Started 182350B system overviewBefore you install Check your shipmentWindows XP Service Pack 3 or later System requirements1Ghz 32- bit x86, 1GHz 64- bit x64, no support for Itanium64 Create an emergency repair disk Installing the Agilent IO Libraries Suite Checking for installed Agilent IO LibrariesInstalling and Configuring Installing Agilent IO Libraries Suite Click Start Run Installing and Configuring the 82350 If you accept the Typical Installation, click Next and pro Installing and Configuring the 82350 When you are done selecting features, click Next Installing and Configuring the 82350 T E Connecting your instruments Installing the 82350 in your PC4Remove the PC cover Connecting instruments to Installing and Configuring 8Connecting a single Gpib instrument Example Connecting Multiple Gpib Instruments Configuring Installing configuration filesConfiguring the 82350 interface 10Agilent Connection Experts11Configure Gpib card parameters System Installing and Configuring the 82350 Communicating with your instruments Verifying instrument communicationProgramming your instruments 12Agilent Interactive IOAccessing Programming Manuals and Help Example Gpib Interface Configuration 13Interface configuration for multiple instruments Troubleshooting Guidelines Troubleshooting Overview 1Troubleshooting overviewCheck cables/connections/power Hardware ChecksCheck Device Manager Number ffffffff Troubleshooting Guidelines 82350B Installation and Configuration Guide Disable Connection Expert Auto-Discovery Software Checks82350B Installation and Configuration Guide Set 82350 Read/Write Performance Mode 82350B Installation and Configuration Guide Troubleshooting Guidelines Check IO Libraries Suite installation Agilent IO Libraries Suite ChecksCheck IO Control operation Install IO Libraries Suite if 82350 was installed first Windows XP/Vista/7 Specifications Information General Requirements General Characteristics +5V PCIAccessing an Electronic Copy of this Guide Contacting Agilent URLSpecifications and Information Index GpibWeb sites, Agilent Contact us