Agilent Technologies 82350B manual Communicating with your instruments

Page 40

1 Installing and Configuring the 82350

Step 5: Communicating with your instruments

After the 82350 has been configured and you have connected your GPIB instruments to the 82350, the next step is to verify communication between your PC and the instruments using the Interactive IO utility. After communication has been established, you can begin programming the instruments using VISA, VISA COM, SICL, or SCPI commands. This section includes:

Verifying Instruments

Programming Your Instruments (Optional)

Verifying instrument communication

N O T E

When the Agilent IO Libraries Suite was installed on your PC, an I/O utility called Interactive IO was also installed. You can use Interactive IO to verify communication between your PC and connected GPIB instruments.This section shows one way to use Interactive IO to verify instrument communication.

Once your GPIB interface has been configured in Connection Expert, if you can see the attached GPIB instrument(s) in the Connection Expert explorer and see their IDN string information in the detail pane then communication has been verified. Interactive IO allows you to manually verify communication and send specific commands to your instruments.

To use Interactive IO to send a *IDN? (identification) query to an instrument:

1Select the instrument by clicking its icon in the Connection Expert explorer view.

2Right- click the instrument and click Send Commands To This Instrument to display the Interactive IO window. For information on

28

82350B Installation and Configuration Guide

Image 40
Contents PCI Gpib Manual Part Number Safety Notices Safety Symbols Page General Safety Information Environment Conditions Regulatory Markings Page Declaration of Conformity Contents Specifications and Information Installing and Configuring Getting Started 182350B system overviewBefore you install Check your shipmentWindows XP Service Pack 3 or later System requirements1Ghz 32- bit x86, 1GHz 64- bit x64, no support for Itanium64 Create an emergency repair disk Installing the Agilent IO Libraries Suite Checking for installed Agilent IO LibrariesInstalling and Configuring Installing Agilent IO Libraries Suite Click Start Run Installing and Configuring the 82350 If you accept the Typical Installation, click Next and pro Installing and Configuring the 82350 When you are done selecting features, click Next Installing and Configuring the 82350 T E Connecting your instruments Installing the 82350 in your PC4Remove the PC cover Connecting instruments to Installing and Configuring 8Connecting a single Gpib instrument Example Connecting Multiple Gpib Instruments Configuring Installing configuration filesConfiguring the 82350 interface 10Agilent Connection Experts11Configure Gpib card parameters System Installing and Configuring the 82350 Communicating with your instruments Verifying instrument communicationProgramming your instruments 12Agilent Interactive IOAccessing Programming Manuals and Help Example Gpib Interface Configuration 13Interface configuration for multiple instruments Troubleshooting Guidelines Troubleshooting Overview 1Troubleshooting overviewCheck cables/connections/power Hardware ChecksCheck Device Manager Number ffffffff Troubleshooting Guidelines 82350B Installation and Configuration Guide Disable Connection Expert Auto-Discovery Software Checks82350B Installation and Configuration Guide Set 82350 Read/Write Performance Mode 82350B Installation and Configuration Guide Troubleshooting Guidelines Check IO Libraries Suite installation Agilent IO Libraries Suite ChecksCheck IO Control operation Install IO Libraries Suite if 82350 was installed first Windows XP/Vista/7 Specifications Information General Requirements General Characteristics +5V PCIAccessing an Electronic Copy of this Guide Contacting Agilent URLSpecifications and Information Index GpibWeb sites, Agilent Contact us