Barco MGD 521 MKII user manual Signal connection, Connection of video signals

Page 13

2.4Signal connection

a)Connection of video signals

!The connected equipment must comply to all relevant safety demands.

Important: The use of low-quality video cables can distort the video signal and influence diagnosis.

1.Check the impedance of the imaging board that produces the video signals you want to connect. It must be 50 or 75 Ohm (just like the input of the display). If not, the quality of the images on the display will be inferior.

2.Connect the video output of the computer’s imaging board to the video and sync inputs on the display’s rear panel. Use a proper video cable. The video cable is not supplied with the display, unless the display comes as part of a complete BARCO MeDis system, that also contains an imaging board.

The inputs accept the following signals:

Video with separate horizontal and vertical sync.

-The video cable has 3 wires.

-Connect the video signal to the connector Video {7}.

-Connect the horizontal sync signal to the connector HS/CS {8}.

-Connect the vertical sync signal to the connector VS {9}.

Video with external composite sync.

-The video cable has 2 wires.

-Connect the video signal to the connector Video {7}.

-Connect the composite sync signal to the connector HS/CS {8}.

Video with internal composite sync (sync on video).

-The video cable has 1 wire.

-Connect the video (with sync) signal to the connector Video {7}.

-The vertical sync pulse must be at least 2 lines long.

Notes:

The video inputs cannot be connected in loop-through (daisy- chain).

The required video amplitude: 700 mV.

The required sync. amplitude: 500 mV.

Installation 14

Image 13
Contents V599326/Rev June MGD 521 MkiiTrademark information MGD 521 M Mkii Users Guide Safety Instructions MGD 521 M Mkii Users Guide Contents MGD 521 M Mkii Users Guide Overview IntroductionVersions and options Power saving systemAbout the manuals Precautions InstallationPackage contents Controls and connectors Connection of video signals Signal connectionConnection of optical sensor Connection of data signalsPositioning the display Connection of powerImportant considerations Power on/off Switching on / offAutomatic power saving system Dpms Operation User ControlsTo activate the OSD Contrast and brightness controlTo put contrast in calibrated position To put brightness in calibrated position To control contrast manuallyTo control brightness manually To view the status information Viewing Status informationExplanation of the Status information menu items Activity timeout Other adjustmentsSettings Advanced Settings and Adjustments For Service Staff onlyTo change the settings Saving the settings Explanation of the Settings menu itemsHow to make the adjustments AdjustmentsDescription of the adjustments Geometry RotationAdvanced Interlacing Post correction Interlacing Pre correctionSH Compensation optional SH Compensation SH Compensation ON/OFFSH Compensation SH Position Advanced Advanced Focus Static focus To adjust the ALC system, proceed as followsFocus zone ALC optionalHow to save changes Proceed as follows Factory resetLuminance uniformity What happens upon saving?Picture tube MaintenanceCabinet There appears no image, the green LED is on TroubleshootingLuminance Uniformity Correction Appendix a Background InformationPage Picture tube Appendix B Technical Specifications