Teledyne 3190 manual Unpacking the Analyzer, Trace Oxygen Analyzer Installation

Page 19

Trace Oxygen Analyzer

Installation 3

 

 

 

 

 

 

 

 

 

Installation

Overvoltage Category II

Installation of the analyzer includes:

1.Unpacking the system.

2.Mounting the Control Unit, External Sample Block, and Micro- Fuel Cell sensor.

3.Making the electrical connections.

5.Making the gas connections.

6.Testing the installation.

CAUTIONS: Read this chapter in its entirety before installing the units.

The Model 3190 is for or indoor use only.

The sample must be free of entrained solids or water. However, a high humidity sample is ideal, since it will prevent water loss from the cell electrolyte.

The Micro-Fuel Cell sensor electrolyte is caustic. Do not attempt to open it. Leaking or exhausted cells should be disposed of in accordance with local regulations. Refer to the Material Safety Data Sheet in the Appendix.

Any damage or scarring of the delicate permeable membrane on the sensing end of the cell will require cell replacement. Prevent contact with membrane by any solid object.

3.1Unpacking the Analyzer

As soon as you receive the instrument, carefully unpack and inspect Control Unit, External Probe, and any included accessories for damage. Immediately report any damage to the shipping agent. The analyzer is

Teledyne Analytical Instruments

3-1

Image 19
Contents Teledyne Analytical Instruments Model 3190 SeriesCopyright 1999 Teledyne Analytical Instruments Specific Model Information Contents Appendix MaintenanceTeledyne Analytical Instruments Trace Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewIntroduction Model Front Panel DescriptionUp Arrow Rear Panel DescriptionRear Panel AC and DC versions Analog OutputsRS-232 Port Introduction Model Teledyne Analytical Instruments Trace Oxygen Analyzer Operational Theory Operational TheoryIntroduction Micro-Fuel Cell SensorAnatomy of a Micro-Fuel Cell Operational Theory Model+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsElectronics Signal ProcessingGeneral Block Diagram of the Signal Processing Electronics Trace Oxygen Analyzer Installation InstallationUnpacking the Analyzer Location and Mounting Installation ModelControl Unit Installation External Probe InstallationElectrical Connections Installing the Micro-Fuel Cell / Cell Block OrientationRear Panel Electrical Connectors for AC and DC Units Trace Oxygen Analyzer Installation Sensor Fail HI AlarmLO Alarm Parameter SettingGas Connections Installation ChecklistInstallation Model Teledyne Analytical Instruments Operation Using the Function and Data EntryTrace Oxygen Analyzer Operation Buttons Setting the Analysis RangesOperation Model Trace Oxygen Analyzer Operation HI Range Setting the Alarm SetpointsSettle Mode HI AlarmCalibration LO AlarmSensor Fail Alarm Selecting a Fixed Range or AutorangingDisplaying Percent & PPM on the LED Display ProcedureSetL mode in the LED display Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance AC Powered UnitsWhen to Replace a Sensor Sensor Installation or ReplacementMaintenance Model DC Powered UnitsRemoving the Micro-Fuel Cell Disconnect the Power Source at the Control UnitExploded View of MFC and Cell Block Installing a Micro-Fuel CellCell Warranty Conditions Disconnect the Power Source from the Control UnitMaintenance Model Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model Spare Parts ListMiscellaneous Drawing ListAppendix Model Teledyne Analytical Instruments