3 Installation | Model 3190 | |
|
|
|
|
|
|
|
|
|
Figure | |
The specific descriptions for each type of alarm are as follows: | |
HI Alarm | Configured as high alarm (actuates when concentration is |
| above threshold). Can be set anywhere within the full |
| range of the analyzer |
| ABOVE the threshold set for the LO Alarm. |
LO Alarm | Configured as low alarm (actuates when concentration is |
| below threshold). Can be set anywhere within the full |
| range of the analyzer |
| BELOW the threshold set for the HI Alarm. |
Sensor Fail | Actuates when the output of the |
| falls below the acceptable level. |
The
Table
Parameter | Setting |
Baud | 2400 |
Byte | 8 bits |
Parity | none |
Stop Bits | 1 |
Message Rate | 2 per second |
External Probe: The receptacle for the analysis unit cable is located in the lower
Teledyne Analytical Instruments |