Teledyne 3190 Replacing the Fuse, Trace Oxygen Analyzer Maintenance, Overvoltage Category

Page 33

Trace Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

Overvoltage Category II

Maintenance

Aside from normal cleaning and checking for leaks, the Model 3190 should not require any maintenance beyond replacement of expended Micro- Fuel Cells, and perhaps a blown fuse. Routine maintenance includes occa- sional recalibration, as described in chapter 4, Operation.

5.1Replacing the Fuse

5.1.1 AC Powered Units

When a fuse blows, check first to determine the cause, then replace the fuse using the following procedure:

1.Disconnect the AC power and place the power switch located on the rear panel in the O position. Remove the power cord from the receptacle.

2.The fuse receptacle is located in the power cord receptacle assembly in the upper left-hand corner of the rear panel. See Figure 5-1.

Figure 5-1: AC Fuse Replacement

3.Insert a small flat-blade screwdriver into the slot in the receptacle wall nearest the fuse and gently pry open the fuse receptacle. The fuse holder will slide out. The fuse in use is visible in the clip. To

Teledyne Analytical Instruments

5-1

Image 33
Contents Teledyne Analytical Instruments Model 3190 SeriesCopyright 1999 Teledyne Analytical Instruments Specific Model Information Contents Appendix MaintenanceTeledyne Analytical Instruments Introduction Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionIntroduction Model Front Panel DescriptionUp Arrow Rear Panel DescriptionRear Panel AC and DC versions Analog OutputsRS-232 Port Introduction Model Teledyne Analytical Instruments Introduction Operational TheoryMicro-Fuel Cell Sensor Trace Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model+ 2H 2O + 4e → 4OH Electrochemical ReactionsEffect of Pressure Calibration CharacteristicsSignal Processing ElectronicsGeneral Block Diagram of the Signal Processing Electronics Installation Trace Oxygen Analyzer InstallationUnpacking the Analyzer Control Unit Installation Installation ModelExternal Probe Installation Location and MountingElectrical Connections Installing the Micro-Fuel Cell / Cell Block OrientationRear Panel Electrical Connectors for AC and DC Units Trace Oxygen Analyzer Installation LO Alarm HI AlarmParameter Setting Sensor FailGas Connections Installation ChecklistInstallation Model Teledyne Analytical Instruments Using the Function and Data Entry OperationTrace Oxygen Analyzer Operation Setting the Analysis Ranges ButtonsOperation Model Settle Mode Setting the Alarm SetpointsHI Alarm Trace Oxygen Analyzer Operation HI RangeSensor Fail Alarm LO AlarmSelecting a Fixed Range or Autoranging CalibrationDisplaying Percent & PPM on the LED Display ProcedureSetL mode in the LED display Trace Oxygen Analyzer Maintenance MaintenanceAC Powered Units Replacing the FuseMaintenance Model Sensor Installation or ReplacementDC Powered Units When to Replace a SensorRemoving the Micro-Fuel Cell Disconnect the Power Source at the Control UnitExploded View of MFC and Cell Block Installing a Micro-Fuel CellCell Warranty Conditions Disconnect the Power Source from the Control UnitMaintenance Model Teledyne Analytical Instruments Trace Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model Spare Parts ListMiscellaneous Drawing ListAppendix Model Teledyne Analytical Instruments